Clinical reference dosimetry: Comparison between AAPM TG-21 and TG-51 protocols
- 6 June 2000
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 27 (6) , 1217-1225
- https://doi.org/10.1118/1.598999
Abstract
We compare the results of absorbed dose determined at reference conditions according to the AAPM TG-21 dose calibration protocol and the new AAPM TG-51 protocol. The AAPM TG-21 protocol for absorbed dose calibration is based on ionization chambers having exposure calibration factors for 60Co gamma rays, N(x). The new AAPM TG-51 dosimetry protocol for absorbed dose calibration is based on ionization chambers having 60Co absorbed dose-to-water calibration factor, N60Co(D,w). This study shows that the dose changes are within 1% for a cobalt beam, 0.5% for photon energies of 6 and 18 MV, and 2%-3% for electron beams with energies of 6 to 20 MeV. The chamber primary calibration factors, Nx and N60Co(D,w), are traceable to the Canadian primary standards laboratory (NRCC). We also present estimated dose changes between the two protocols when calibration factors are traceable to NIST in the United States.Keywords
This publication has 15 references indexed in Scilit:
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Measured overall perturbation factors at depths greater than for ionization chambers in electron beamsMedical Physics, 1999
- Measurement of Prepl Pwall factors in electron beams and in a 60Co beam for plane-parallel chambersMedical Physics, 1998
- A new approach to electron‐beam reference dosimetryMedical Physics, 1998
- Experimental determination of fluence correction factors at depths beyond dmax for a Farmer type cylindrical ionization chamber in clinical electron beamsMedical Physics, 1997
- R50 as a beam quality specifier for selecting stopping‐power ratios and reference depths for electron dosimetryMedical Physics, 1996
- The calibration and use of plane‐parallel ionization chambers for dosimetry of electron beams: An extension of the 1983 AAPM protocol report of AAPM Radiation Therapy Committee Task Group No. 39Medical Physics, 1994
- Verification of absorbed doses determined with thimble and parallel-plate ionization chambers in clinical electron beams using ferrous sulphate dosimetryMedical Physics, 1994
- Clinical electron‐beam dosimetry: Report of AAPM Radiation Therapy Committee Task Group No. 25Medical Physics, 1991
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983