Improved recovery time in single coil pulsed NMR spectrometers
- 1 October 1975
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 46 (10) , 1342-1343
- https://doi.org/10.1063/1.1134069
Abstract
A simple modification to a conventional single coil pulsed NMR spectrometer is described which damps the final stage of the transmitter at low rf levels. For an unmodified Bruker SXP spectrometer a receiver recovery time of 7.7 μsec is obtained at 60 MHz; this time improves to 5.5 μsec on adding the ’’low level damper.’’Keywords