Abstract
Accurate determination of relaxation times has become increasingly important in efforts to determine the diagnostic specificity of NMR imaging. Techniques used in NMR imaging, not routinely employed in conventional NMR spectroscopy, can significantly affect the resulting relaxation time determinations. For the saturation recovery (SR) approach of T1 measurement, these include selective excitation to define the image plane and magnetization refocusing for NMR signal acquisition. Computer modeling of the Bloch equations shows that errors well over 50% can be made in image derived T1 measurements if the conventional SR relation between signal intensity and the 90.degree.-90.degree. interpulse delay, tr, is used. Corrected expressions can be derived for the actual pulse and gradient sequence used and phantom data acquired in imaging experiments have verified the validity of these equations. This allows for the correction of T1 data to compensate for systematic bias introduced during imaging procedures and suggests a means whereby errors introduced by radio frequency inhomogeneities across the imaging volume can be reduced.This publication has 4 references indexed in Scilit:
- NMR imaging of leg tumors.Radiology, 1983
- Nuclear magnetic resonance whole-body imager operating at 3.5 KGauss.Radiology, 1982
- Nuclear magnetic resonance transverse relaxation in muscle waterBiophysical Journal, 1981
- PROLONGATION OF PROTON SPIN-LATTICE RELAXATION-TIMES IN REGIONALLY ISCHEMIC TISSUE FROM DOG HEARTS1980