Mass calibration of a matrix‐assisted laser desorption/ionization time‐of‐flight mass spectrometer including the rise time of the delayed extraction pulse
- 30 December 2002
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 17 (3) , 229-237
- https://doi.org/10.1002/rcm.898
Abstract
To utilize fully modern MALDI-TOF and TOF/TOF mass spectrometers with mass resolution exceeding 10 000 and 2 ppm precision of flight time measurements for high mass accuracy, the model of ion motion used in the mass calibration equation must be expanded. The standard three-term equation providing up to 5–10 ppm (rms) mass accuracy with internal standards was modified with an additional term accounting for the finite rise time of the high-voltage extraction pulse. This new four-term calibration equation minimizes the effect of systematic error resulting from the fact that ion velocities are mass dependent due to the rise time of the extraction pulse. Applying this new calibration equation to a mass spectrum obtained in an axial MALDI-TOF MS containing 70 peaks (sodiated PEG), each with a signal-to-noise ratio greater than 100, a mass accuracy of 1.6 ppm (rms) was obtained over the mass range 1.0–4.0 kDa compared with 3.6 ppm (rms) with the standard three-term equation. The physical basis of the effects of the finite extraction pulse rise time on mass calibration is examined for axial MALDI-TOF mass spectrometers, as well as for orthogonal acceleration TOF mass spectrometers. Copyright © 2002 John Wiley & Sons, Ltd.Keywords
This publication has 23 references indexed in Scilit:
- Measurements of mean initial velocities of analyte and matrix ions in infrared matrix-assisted laser desorption ionization mass spectrometryJournal of the American Society for Mass Spectrometry, 2002
- Delayed extraction experiments using a repulsive potential before ion extraction: evidence of clusters as ion precursors in UV-MALDI. Part I: dynamical effects with the matrix 2,5-dihydroxybenzoic acidInternational Journal of Mass Spectrometry, 2002
- Peer Reviewed: Orthogonal-Injection TOFMS for Analyzing Biomolecules.Analytical Chemistry, 1999
- Peptide and protein identification by matrix-assisted laser desorption ionization (MALDI) and MALDI-post-source decay time-of-flight mass spectrometryJournal of the American Society for Mass Spectrometry, 1999
- Resolution and mass accuracy in matrix-assisted laser desorption ionization-time-of-flightJournal of the American Society for Mass Spectrometry, 1998
- Protein indentification using mass spectrometric informationElectrophoresis, 1998
- Delayed extraction matrix‐assisted laser desorption time‐of‐flight mass spectrometryRapid Communications in Mass Spectrometry, 1995
- Error-Tolerant Identification of Peptides in Sequence Databases by Peptide Sequence TagsAnalytical Chemistry, 1994
- Improved Resolution and Very High Sensitivity in MALDI TOF of Matrix Surfaces Made by Fast EvaporationAnalytical Chemistry, 1994
- Velocity distributions of intact high mass polypeptide molecule ions produced by matrix assisted laser desorptionChemical Physics Letters, 1991