Fat quantification with IDEAL gradient echo imaging: Correction of bias from T1 and noise
Top Cited Papers
Open Access
- 24 July 2007
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 58 (2) , 354-364
- https://doi.org/10.1002/mrm.21301
Abstract
Quantification of hepatic steatosis is a significant unmet need for the diagnosis and treatment of patients with nonalcoholic fatty liver disease (NAFLD). MRI is capable of separating water and fat signals in order to quantify fatty infiltration of the liver (hepatic steatosis). Unfortunately, fat signal has confounding T1 effects and the nonzero mean noise in low signal-to-noise ratio (SNR) magnitude images can lead to incorrect estimation of the true lipid percentage. In this study, the effects of bias from T1 effects and image noise were investigated. An oil/water phantom with volume fat-fractions ranging linearly from 0% to 100% was designed and validated using a spoiled gradient echo (SPGR) sequence in combination with a chemical-shift based fat-water separation method known as iterative decomposition of water and fat with echo asymmetry and least squares estimation (IDEAL). We demonstrated two approaches to reduce the effects of T1: small flip angle (flip angle) and dual flip angle methods. Both methods were shown to effectively minimize deviation of the measured fat-fraction from its true value. We also demonstrated two methods to reduce noise bias: magnitude discrimination and phase-constrained reconstruction. Both methods were shown to reduce this noise bias effectively from 15% to less than 1%. Magn Reson Med 58:354–364, 2007.Keywords
This publication has 34 references indexed in Scilit:
- Wilson's disease: an updateNature Clinical Practice Neurology, 2006
- Single acquisition water‐fat separation: Feasibility study for dynamic imagingMagnetic Resonance in Medicine, 2005
- Iterative decomposition of water and fat with echo asymmetry and least‐squares estimation (IDEAL): Application with fast spin‐echo imagingMagnetic Resonance in Medicine, 2005
- Cramér–Rao bounds for three‐point decomposition of water and fatMagnetic Resonance in Medicine, 2005
- Multicoil Dixon chemical species separation with an iterative least‐squares estimation methodMagnetic Resonance in Medicine, 2003
- Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady stateMagnetic Resonance in Medicine, 2003
- Fatty liver in familial hypobetalipoproteinemia: triglyceride assembly into VLDL particles is affected by the extent of hepatic steatosisJournal of Lipid Research, 2003
- Imaging of Hepatic SteatosisSeminars in Liver Disease, 2001
- Comparison of In-Phase and Opposed-Phase GRE and Conventional SE MR Pulse Sequences in T1-Weighted Imaging of Liver LesionsJournal of Computer Assisted Tomography, 1996
- Fatty liver hepatitis (steatohepatitis) and obesity: An autopsy study with analysis of risk factorsHepatology, 1990