A comparison of linear and orthogonal regression analysis for local INR determination in ECAA coagulometer studies
- 1 September 1998
- journal article
- research article
- Published by Wiley in British Journal of Haematology
- Vol. 102 (4) , 910-917
- https://doi.org/10.1046/j.1365-2141.1998.00866.x
Abstract
International sensitivity index calibrations based on the W.H.O. recommended method depend on orthogonal regression analysis. As this is not readily available in statistical packages, comparison has been made with simple linear regression analysis in a study of coagulometer effects on the International Normalized Ratio (INR) at 155 European centres. Sets of seven lyophilized normal and 20 lyophilized artificially depleted abnormal plasmas were provided with five coumarin test plasmas and two European Concerted Action on Anticoagulation reference thromboplastins (low International Sensitivity Index (ISI) human and high ISI rabbit). Local ISI based on the artificially depleted lyophilized plasmas using conventional orthogonal regression gave good correction for local coagulometer effects on the human reagent and minimal correction with the rabbit reagent INR. Results were considerably worse after attempts at correction using calibration based on linear regression analysis with both reagents. The results indicate that calibration of coagulometer prothrombin time systems using simple linear regression is not appropriate.Keywords
This publication has 21 references indexed in Scilit:
- A comparison of artificially‐depleted, lyophilized coumarin and fresh coumarin plasmas in thromboplastin calibrationBritish Journal of Haematology, 1998
- Local INR correction: justification for a simplified approach.Journal of Clinical Pathology, 1997
- A simplified statistical method for local INR using linear regressionBritish Journal of Haematology, 1997
- System ISI calibration: a universally applicable scheme is possible only when coumarin plasma calibrants are usedBritish Journal of Haematology, 1997
- A partial factor V deficiency in anticoagulated lyophilized plasmas has been identified as a cause of the international normalized ratio discrepancy in the external quality assessment schemeThrombosis Research, 1995
- Assessment of value of calibrated lyophilised plasmas to determine International Sensitivity Index for coagulometers.Journal of Clinical Pathology, 1992
- Comparison of Two Automated Coagulometers and the Manual Tilt-Tube Method for the Determination of Prothrombin TimeAmerican Journal of Clinical Pathology, 1989
- Effect of automation on prothrombin time test in NEQAS surveys.Journal of Clinical Pathology, 1989
- Orthogonal Regression EquationPublished by Springer Nature ,1984
- Gerinnungsphysiologische Schnellmethode zur Bestimmung des FibrinogensActa Haematologica, 1957