Optimum Detection Times for Bacteria and Yeast Species with the BACTEC 9120 Aerobic Blood Culture System: Evaluation for a 5-Year Period in a Turkish University Hospital
Open Access
- 1 February 2003
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 41 (2) , 819-821
- https://doi.org/10.1128/jcm.41.2.819-821.2003
Abstract
We tracked and documented the time of positivity of blood cultures by using the BACTEC 9120 (Becton Dickinson Diagnostic Instrument Systems) blood culture system over a 5-year study period. A 7-day protocol of the incubation period was selected, and a total of 11,156 blood cultures were evaluated. The clinically significant microorganisms (32.95%) were isolated in 3,676 specimens. Gram-positive and -negative bacterial isolation rates were found to be 41.07 and 44.88%, respectively. Yeasts were found in 14.03% of all pathogens. Both the false-positivity and -negativity rates were very low (0.1 and 0.3%, respectively). The mean detection times for all of the pathogens were determined to be 19.45 h. Yeasts, nonfermentative gram-negative bacteria, and Brucella melitensis strains were isolated within 5 days. By taking these data into account, we decided to establish a 5-day-incubation protocol in our laboratory instead of the 7 days that are commonly used.Keywords
This publication has 21 references indexed in Scilit:
- LABORATORY DIAGNOSIS OF BACTEREMIA AND FUNGEMIAInfectious Disease Clinics of North America, 2001
- Use of BACTEC 9240 Blood Culture System for Detection of Brucella melitensis in Synovial FluidJournal of Clinical Microbiology, 2001
- Outpatient Pediatric Blood Cultures: Time to PositivityPediatrics, 2000
- Comparison of Five Days versus Seven Days of Incubation for Detection of Positive Blood Cultures by the Bactec 9240 SystemEuropean Journal of Clinical Microbiology & Infectious Diseases, 1998
- Comparison of five days versus seven days of incubation for detection of positive blood cultures by the Bactec 9240 systemEuropean Journal of Clinical Microbiology & Infectious Diseases, 1998
- Experience of changing between signal and Bactec 9240 blood culture systems in a children's hospital.Journal of Clinical Pathology, 1998
- Multicenter clinical comparison of resincontaining bottles with standard aerobic and Anaerobic bottles for culture of microorganisms from bloodEuropean Journal of Clinical Microbiology & Infectious Diseases, 1997
- Rapid detection ofBrucella melitensis from blood cultures by a commercial systemEuropean Journal of Clinical Microbiology & Infectious Diseases, 1997
- Detection of Brucella melitensis by BACTEC NR 730 and BACTEC 9120 systemsEuropean Journal of Epidemiology, 1996
- The impact of microbiological diagnostics on the antimicrobial treatment of hospitalised patients with infectious disease.1996