Influence of detector size in photon beam profile measurements
- 1 May 1991
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 36 (5) , 621-631
- https://doi.org/10.1088/0031-9155/36/5/005
Abstract
Correction is necessary to account for the detector size in clinical dosimetry of photon and electron beams. This correction is due to the absorbed dose gradient present in a finite-size detector. Further corrections are necessary when the detector and phantom materials are not the same. These corrections are due to the perturbation in the charged-particle fluence. Generally these corrections are applied to measurements along the central axis of the beam. Cross beam profile measurements, however, are not usually corrected for detector size. The ionization profile is also usually assumed to be equivalent to the absorbed dose profile. We have corrected the ionization chamber size effect by two approaches: extrapolation of measurements to zero detector size and deconvolution of measurements using a simple model for the detector response function. We have measured absorbed dose profiles to water using a small water-equivalent plastic scintillation detector. Film profile measurements were also studied. The ionization profile corrected for detector size and absorbed dose profile were not equal, probably due to loss of charged-particle equilibrium in the beam edges. For ionization chamber measurements, knowledge of the charged-particle spectrum is needed to convert ionization to absorbed dose to water. This is not necessary for relative absorbed dose measurements under charged-particle equilibrium. Film has been shown to be a straightforward and reliable method for cross beam profile measurements.Keywords
This publication has 15 references indexed in Scilit:
- Film dosimetry: linearisation of dose-response for relative measurements of dose distributionPhysics in Medicine & Biology, 1990
- Measurements of dose distributions in small beams of 6 MV X-raysPhysics in Medicine & Biology, 1987
- Radiation Dosimetry: Physical and Biological Aspects edited by C. G. OrtonMedical Physics, 1986
- Fluence perturbation in photon beams under nonequilibrium conditionsMedical Physics, 1986
- Penumbral measurements in water for high‐energy x raysMedical Physics, 1986
- A new look at displacement factor and point of measurement corrections in ionization chamber dosimetryMedical Physics, 1983
- Comparison of film and ion chamber systems for depth-dose measurements for a 25 MV beamPhysics in Medicine & Biology, 1979
- Einfluß der Vielfachstreuung von Elektronen auf die Ionisation in gasgefüllten HohlräumenRadiation and Environmental Biophysics, 1968
- Effect of Finite Size of Ionization Chambers on Measurements of Small Photon SourcesRadiation Research, 1960
- Effect of Dosemeter Size on Measurements Close to a Radioactive SourceThe British Journal of Radiology, 1959