Calibration of a Kerr Cell System for High-Voltage Pulse Measurements
- 1 January 1968
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 17 (4) , 313-320
- https://doi.org/10.1109/tim.1968.4313725
Abstract
Several techniques for calibration of an electrooptical (Kerr cell) high-voltage pulse measuring system are described. Independent calibrations, without reference to pulse divider measurements, are achieved by application of a direct bias voltage to the Kerr cell. After calibration, experiments with voltages as high as 100 kV demonstrate reasonable agreement (to within 1 percent) between simultaneous Kerr cell and calibrated pulse divider measurements.Keywords
This publication has 5 references indexed in Scilit:
- Kerr Cell Measuring System for High Voltage PulsesReview of Scientific Instruments, 1964
- High Voltage Pulse Measuring System Based on Kerr EffectReview of Scientific Instruments, 1963
- Spark-gap flashover measurements for steeply rising voltage impulsesJournal of Research of the National Bureau of Standards, Section C: Engineering and Instrumentation, 1962
- An Electro-Optical Shutter for Photographic PurposesTransactions of the American Institute of Electrical Engineers, 1949
- XL. A new relation between electricity and light: Dielectrified media birefringentJournal of Computers in Education, 1875