Comparison of SmartCycler Real-Time Reverse Transcription-PCR Assay in a Public Health Laboratory with Direct Immunofluorescence and Cell Culture Assays in a Medical Center for Detection of Influenza A Virus
Open Access
- 1 August 2003
- journal article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 41 (8) , 3597-3601
- https://doi.org/10.1128/jcm.41.8.3597-3601.2003
Abstract
A single-tube real-time (fluorogenic) reverse transcription (RT)-PCR with the SmartCycler instrument (SmartCycler RT-PCR) for influenza A virus detection was evaluated with 238 respiratory specimens. Direct immunofluorescence antibody staining (DFA) and primary rhesus monkey kidney cell culture were performed on-site at Yale-New Haven Hospital. Specimens were transported to the Connecticut Department of Public Health Laboratory for real-time RT-PCR. Cell culture detected influenza A virus in all 150 influenza A virus-positive specimens, DFA detected the virus in 148 influenza A virus-positive specimens, and SmartCycler RT-PCR detected the virus 143 influenza A virus-positive specimens. The sensitivity and specificity of RT-PCR were 95.3 and 100%, respectively. The high sensitivity and specificity and the rapid turnaround time made the SmartCycler RT-PCR valuable for the rapid diagnosis of influenza A, especially in a public health laboratory. The closed real-time RT-PCR system avoided cross-contamination possible with RT-PCR and the excessive manipulations required for conventional RT-PCR analysis and saved time and labor as well. In a medical center, rapid diagnosis by DFA was labor intensive but was 98.7% sensitive and 100% specific compared to the results of culture and provided results within 2 h throughout operating hours, helping with bed allocation on admission and patient management.Keywords
This publication has 31 references indexed in Scilit:
- Suboptimal Detection of Influenza Virus in Adults by the Directigen Flu A+B Enzyme Immunoassay and Correlation of Results with the Number of Antigen-Positive Cells Detected by Cytospin ImmunofluorescenceJournal of Clinical Microbiology, 2003
- Rapid Enterovirus RNA Detection in Clinical Specimens by Using Nucleic Acid Sequence-Based AmplificationJournal of Clinical Microbiology, 2003
- Practical and Sensitive Screening Strategy for Detection of Influenza VirusJournal of Clinical Microbiology, 2002
- Evaluation of a New Dot Blot Enzyme Immunoassay (Directigen Flu A+B) for Simultaneous and Differential Detection of Influenza A and B Virus Antigens from Respiratory SamplesJournal of Clinical Microbiology, 2002
- Effectiveness of Reverse Transcription-PCR, Virus Isolation, and Enzyme-Linked Immunosorbent Assay for Diagnosis of Influenza A Virus Infection in Different Age GroupsJournal of Clinical Microbiology, 2002
- Evaluation of the Directigen FluA+B Test for Rapid Diagnosis of Influenza Virus Type A and B InfectionsJournal of Clinical Microbiology, 2002
- Polymerase Chain Reaction Is More Sensitive than Viral Culture and Antigen Testing for the Detection of Respiratory Viruses in Adults with Hematological Cancer and PneumoniaClinical Infectious Diseases, 2002
- Evaluation of a single‐step multiplex RT‐PCR for influenza virus type and subtype detection in respiratory samplesJournal of Clinical Laboratory Analysis, 2002
- Evaluation of a Rapid Optical Immunoassay for Influenza Viruses (FLU OIA Test) in Comparison with Cell Culture and Reverse Transcription-PCRJournal of Clinical Microbiology, 2001
- Simultaneous Detection of Influenza Viruses A and B Using Real-Time Quantitative PCRJournal of Clinical Microbiology, 2001