Bubble CPAP: Is the Noise Important? An In Vitro Study
- 1 June 2005
- journal article
- Published by Springer Nature in Pediatric Research
- Vol. 57 (6) , 826-830
- https://doi.org/10.1203/01.pdr.0000157721.66812.07
Abstract
Continuous positive airway pressure (CPAP) is often used to provide noninvasive respiratory support in infants with Respiratory distress syndrome. The recruitment of atelectatic lung and appropriate lung volume maintenance are vital to the success of CPAP treatment. The noisy pressure waveform of bubble CPAP superimposed on pressure fluctuations as a result of spontaneous breathing may promote airway opening events as a result of stochastic resonance. The magnitude and the frequency of the superimposed noise are critical to this process. We hypothesized that the applied bias flow and mechanical properties of the lung would influence the magnitude and the frequency content of the noise transmitted to the lung. The effect of varying bias flow (6–10 L/min) and lung compliance (0.1–1.5 mL/cm H2O) on the mean, range, and frequency content of the pressure fluctuations at the airway opening and within the lung was evaluated in an in vitro model lung. Increasing bias flow increased the mean and the magnitude of pressure oscillations at the airway opening and in the lung model. Decreasing compliance of the lung model increased the magnitude and the frequency content of pressure oscillations in the model lung. Lung mechanics and applied flow influence the magnitude of the noise superimposed on the transmitted pressure waveform and may influence lung volume recruitment in bubble CPAP.Keywords
This publication has 21 references indexed in Scilit:
- Acoustic evidence of airway opening during recruitment in excised dog lungsJournal of Applied Physiology, 2004
- Dependence of Intrapulmonary Pressure Amplitudes on Respiratory Mechanics during High-Frequency Oscillatory Ventilation in Preterm LambsPediatric Research, 2002
- Decreased Indicators of Lung Injury with Continuous Positive Expiratory Pressure in Preterm LambsPediatric Research, 2002
- Attenuation of pressure swings along the endotracheal tube is indicative of optimal distending pressure during high‐frequency oscillatory ventilation in a model of acute lung injuryPediatric Pulmonology, 2002
- Variable Tidal Volume Ventilation Improves Lung Mechanics and Gas Exchange in a Rodent Model of Acute Lung InjuryAmerican Journal of Respiratory and Critical Care Medicine, 2002
- In Vitro Performance Characteristics of High-Frequency Oscillatory VentilatorsAmerican Journal of Respiratory and Critical Care Medicine, 2001
- Biologically Variable or Naturally Noisy Mechanical Ventilation Recruits Atelectatic LungAmerican Journal of Respiratory and Critical Care Medicine, 2000
- Do Clinical Markers of Barotrauma and Oxygen Toxicity Explain Interhospital Variation in Rates of Chronic Lung Disease?Pediatrics, 2000
- Avalanches and power-law behaviour in lung inflationNature, 1994
- A new device for CPAP by nasal routeCritical Care Medicine, 1975