High-resolution circular ring positron tomograph with dichotomic sampling: Dichotom-I
- 1 November 1983
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 28 (11) , 1219-1234
- https://doi.org/10.1088/0031-9155/28/11/003
Abstract
The circular ring transaxial positron camera developed earlier (see S.E. Derenzo et al., IEEE Trans. Nucl. Sci., vol. NS-24, p.554-8, 1977) was refitted with a new dichotomic sampling scheme and aperture collimators on the detector array to improve the sampling and the overall system resolution. The z-axis slice thickness collimators were also limited to 1 cm, which corresponds to a slice thickness of 0.5 cm FWHM. Two different types of aperture collimators were adopted for high resolution (HR) and very high resolution (VHR) imaging, respectively. In HR mode a resolution of 6.5 mm FWHM was obtained without appreciable degradation of overall sensitivity, which represents a threefold improvement in resolution over the original system. In phantom studies with HR mode a sensitivity of 4500 counts s-1 mu Ci-1 cm-3 was obtained for a 20 cm diameter uniform phantom filled with water. A VHR mode experiment was also conducted to observe the ultimate resolution capability of the Dichotom-I system, and a resolution of 4.2 mm FWHM was obtained at the expense of sensitivity which was reduced by a factor of four from the HR mode experiment. The experience gained with Dichotom-I suggests a relatively simple and inexpensive modification of the existing NaI(Tl) ring positron cameras, most of which suffer from low resolution due to poor sampling and poor intrinsic detector resolution.Keywords
This publication has 18 references indexed in Scilit:
- Design Considerations for Positron Emission TomographyIEEE Transactions on Biomedical Engineering, 1981
- Effects of Wobbling Motion on Image Quality in Positron TomographyIEEE Transactions on Nuclear Science, 1981
- A New Sampling Scheme for the Ring Positron Camera: Dichotomic Ring SamplingIEEE Transactions on Nuclear Science, 1981
- Design of a High Resolution Positron Emission TomographJournal of Computer Assisted Tomography, 1980
- Sampling Requirements and Detector Motion for Positron Emission TomographyIEEE Transactions on Nuclear Science, 1979
- The Donner 280-Crystal High Resolution Positron TomographIEEE Transactions on Nuclear Science, 1979
- Design and Performance Characteristics of a Production Model Positron Imaging SystemIEEE Transactions on Nuclear Science, 1978
- A Computer Assisted Ringdector Positron Camera System for Reconstruction Tomography of the BrainIEEE Transactions on Nuclear Science, 1978
- Circular Ring Transverse Axial Positron Camera for 3-Dimensional Reconstruction of Radionuclides DistributionIEEE Transactions on Nuclear Science, 1976
- A Multi-Crystal Positron CameraIEEE Transactions on Nuclear Science, 1972