Image quality evaluation of a desktop computed radiography system
- 7 December 2000
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 27 (12) , 2669-2679
- https://doi.org/10.1118/1.1326449
Abstract
The modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) of the Lumisys ACR-2000 desktop computed radiography (CR) reader were measured and compared to equivalent measurements acquired from a Fuji AC-3 CR system. The one-dimensional (1D) MTF was measured from an image of a sharp edge and the 1D NPS was derived from a 2D NPS measured from a uniform field exposure. The energy dependent ideal input signal to noise ratio of the incident x-ray beams was estimated using published x-ray spectra and attenuation coefficients. Measurements were acquired using Agfa, Fuji, and Kodak storage phosphor plates and it was concluded that use of the Fuji plates resulted in the highest system DQE for the ACR-2000. The DQE was measured using exposures of 0.10, 1.0, and 10.0 mR from 70 and 120 kVp x-ray beams filtered with aluminum. The DQE of the Lumisys ACR-2000 was lower than that of the Fuji AC-3.Keywords
This publication has 9 references indexed in Scilit:
- Experimental comparison of noise and resolution for 2k and 4k storage phosphor radiography systemsMedical Physics, 1999
- Performance characteristics of a Kodak computed radiography systemMedical Physics, 1999
- A method for measuring the presampled MTF of digital radiographic systems using an edge test deviceMedical Physics, 1998
- An accurate method for computer‐generating tungsten anode x‐ray spectra from 30 to 140 kVMedical Physics, 1997
- Comparison of x‐ray cross sections for diagnostic and therapeutic medical physicsMedical Physics, 1996
- DQE(f) of four generations of computed radiography acquisition devicesMedical Physics, 1995
- Signal and noise in modulation transfer function determinations using the slit, wire, and edge techniquesMedical Physics, 1992
- Imaging performance of a digital storage phosphor systemMedical Physics, 1987
- A method for modulation transfer function determination from edge profiles with correction for finite-element differentiationMedical Physics, 1987