Strength Recovery in Fractured Sheep Tibia Treated with a Plate or an Internal Fixator: An Experimental Study with a Two-Year Follow-up
- 1 January 1997
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Orthopaedic Trauma
- Vol. 11 (1) , 14-23
- https://doi.org/10.1097/00005131-199701000-00005
Abstract
Comparison of fracture healing with two different implants: a conventional Dynamic Compression Plate (DCP) and a new internal Point Contact Fixator (PC-Fix). Randomized, prospective study in experimental animals. Observation times: 12, 24, 48 and 96 weeks, with six sheep per group. Following surgery, animals were kept with unrestricted weight-bearing in individual stalls for 12 weeks, thereafter in groups. 56 adult Swiss mountain sheep. a standardized oblique fracture of the sheep tibia was reduced and compressed by a lag screw and "neutralized" with one of the implants. Standard radiographs were used for callus size measurements. After sacrifice the implant was removed and both the treated bone and the contralateral bone were tested for static strength in bending with the plate side under tension. Broken bones were processed for histological evaluation. In the DCP group all six bones failed through the original fracture at 12 weeks. At 24 and 48 weeks two out of six, at 96 weeks one out of six bones failed through the original fracture, others through one of the screw holes. In the PC-Fix group there were no failures through the original fracture with a single exception at 96 weeks. The strength values in the PC-Fix groups of 12 and 96 weeks were significantly higher then in the corresponding DCP groups. Healing of simple diaphysial fractures treated by PC-Fix was superior to that achieved by conventional plating. The histological evaluation suggested that the observed differences can be accounted for by the absence of implant-related cortical necrosis and by the circumferentially uninterrupted (if smaller) callus in the PC-Fix group.Keywords
This publication has 7 references indexed in Scilit:
- A Mechanical Comparison of the Dynamic Compression Plate, Limited Contact-Dynamic Compression Plate, and Point Contact FixatorJournal of Orthopaedic Trauma, 1995
- Internal Fixation with a Self-Compressing Plate and Lag ScrewJournal of Orthopaedic Trauma, 1991
- Wave plate osteosynthesis as a salvage procedureArchives of orthopaedic and trauma surgery, 1990
- The mechanical and morphological properties of bone beneath internal fixation plates of differing rigidityJournal of Orthopaedic Research, 1989
- Early Temporary Porosis of Bone Induced by Internal Fixation Implants A Reaction to Necrosis, Not to Stress Protection?Published by Wolters Kluwer Health ,1988
- Some effects of rigidity of internal fixation on the healing pattern of osteotomiesInjury, 1969