An investigation of methods of pileup rejection for 2-D array detectors employed in high resolution PET
- 1 June 1991
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 10 (2) , 223-227
- https://doi.org/10.1109/42.79481
Abstract
Two methods of rejecting pileup events in 2-D array detectors used in positron emission tomography (PET) are investigated. Since the methods are based on different physical properties, they were investigated separately and in combination. Method one is based on the fact that valid events in the 2-D array detector occur only in a limited pulseheight range for each detector element, and by rejecting events outside these energy ranges, many of the pileup events can be rejected. Method two is based on the fact that most pileup event pairs have a time differential, which will cause a difference in the shape of the signal which increases the rise time of an integrated signal or the crossover time of a bipolar signal from a shaping amplifier. Method two proved to be somewhat better than method one. However, the ideal system would most likely use both methods simultaneously.Keywords
This publication has 4 references indexed in Scilit:
- Detector, shielding and geometric design factors for a high-resolution PET systemIEEE Transactions on Nuclear Science, 1990
- A study of data loss and mispositioning due to pileup in 2-D detectors in PETIEEE Transactions on Nuclear Science, 1990
- An evaluation of a two-dimensional array detector for high resolution PETIEEE Transactions on Medical Imaging, 1988
- A preliminary evaluation of a positron camera system using weighted decoding of individual crystalsIEEE Transactions on Nuclear Science, 1988