Pulse oximeter probes A comparison between finger, nose, ear and forehead probes under conditions of poor perfusion*
- 1 April 1991
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 46 (4) , 260-265
- https://doi.org/10.1111/j.1365-2044.1991.tb11492.x
Abstract
The performances of 10 pulse oximeters using finger probes were compared with the same pulse oximeters using alternative probes (eight finger probes, two nose probes and a forehead probe) in poorly perfused patients. All readings were then compared with directly measured arterial blood oxygen saturations. The mean difference (bias, 'accuracy'), standard deviation (precision) and 'drop out' rate for each pulse oximeter combination was determined. An overall ranking of performance of each pulse oximeter was calculated using five criteria (accuracy, precision, number of readings within 3% of standard, percentage of readings given within 3% of standard, expected overread limit in 95% of cases). Nose and forehead probes performed poorly. Some ear probes performed well compared to some finger probes, but the overall performance of probes in other sites compared to finger probes was worse, (p = 0.05). Two of eight ear probes and no nose or forehead probes would be expected to be within 4% of the reference value in 95% of readings. The use of finger probes rather than probes in other sites is recommended in the patient with poor peripheral perfusion.Keywords
This publication has 13 references indexed in Scilit:
- A comparison of the performance of 20 pulse oximeters under conditions of poor perfusionAnaesthesia, 1991
- Principles of Pulse OximetryAnesthesia & Analgesia, 1989
- Peripheral perfusion after cardiac surgeryCritical Care Medicine, 1989
- Pulse oximetryJournal of Clinical Monitoring and Computing, 1989
- Pulse OximetryAnesthesiology, 1989
- Accuracy of Response of Six Pulse Oximeters to Profound HypoxiaAnesthesiology, 1987
- Evaluation of the Ohmeda 3700 Pulse OximeterAnesthesiology, 1987
- ACCURACY OF A PULSE OXIMETER IN THE CRITICALLY ILL ADULTAnesthesiology, 1985
- Noninvasive Detection of Profound Arterial Desaturations Using a Pulse Oximetry DeviceAnesthesiology, 1985
- Evaluation of Pulse OximetryAnesthesiology, 1983