Abstract
In this multicenter evaluation, the VERSANT HCV RNA 3.0 Assay (bDNA) (Bayer Diagnostics, Tarrytown, N.Y.) was shown to have excellent reproducibility, linearity, and analytical sensitivity across specimen collection matrices (serum, EDTA, ACD-A), and hepatitis C virus (HCV) genotypes 1 to 6. The VERSANT HCV bDNA Assay has a reportable range of 615 to 7,690,000 (7.69 × 106) IU/ml. The total coefficient of variation (CV) ranged from 32.4% at 615 IU/ml to 17% at 6.8 × 106IU/ml. The assay was linear across the reportable range. Analytical specificity of 98.8% was determined by testing 999 specimens from volunteer blood donors. Evaluation of HCV genotypes using RNA transcripts of representative clones of 1a, 1b, 2a, 2b, 2c, 3a, 4a, 5a, and 6a and patient specimens showed that the largest difference between genotype 1, upon which the assay is standardized, and non-1 genotypes was within 1.5-fold. Testing of potentially interfering endogenous substances and exogenous substances and conditions found no interference in HCV-positive or HCV-negative specimens except for unconjugated bilirubin at concentrations of ≥20 mg/dl and protein at concentrations of ≥9 g/dl. Biological variability was estimated from 29 clinically stable individuals not on HCV therapy who were tested weekly over an 8-week period. The combined estimate of total (biologic plus assay) variability was 0.15 log10standard deviation (CV, 36.1%), a fold change of 2.6. Thus, the observed fold change between any two consecutive HCV RNA measures is expected to be less than 2.6-fold (equivalent to 0.41 log10IU/ml) 95% of the time in clinically stable individuals.Keywords
This publication has 28 references indexed in Scilit:
- Use and interpretation of virological tests for hepatitis CHepatology, 2002
- Comparison of Methodologies for Quantification of Hepatitis C Virus (HCV) RNA in Patients Coinfected with HCV and Human Immunodeficiency VirusClinical Infectious Diseases, 2002
- Evaluation of the VERSANT HCV RNA 3.0 Assay for Quantification of Hepatitis C Virus RNA in SerumJournal of Clinical Microbiology, 2002
- Performance of the New Bayer VERSANT HCV RNA 3.0 Assay for Quantitation of Hepatitis C Virus RNA in Plasma and Serum: Conversion to International Units and Comparison with the Roche COBAS Amplicor HCV Monitor, Version 2.0, AssayJournal of Clinical Microbiology, 2002
- Calibration of HCV Working Reagents for NAT Assays against the HCV International StandardVox Sanguinis, 2000
- Is an “à la carte” combination interferon alfa-2b plus ribavirin regimen possible for the first line treatment in patients with chronic hepatitis C?Hepatology, 2000
- The Prevalence of Hepatitis C Virus Infection in the United States, 1988 through 1994New England Journal of Medicine, 1999
- Establishment of the First International Standard for Nucleic Acid Amplification Technology (NAT) Assays for HCV RNAVox Sanguinis, 1999
- Hepatitis C: The clinical spectrum of diseaseHepatology, 1997
- Therapy of hepatitis C: Meta-analysis of interferon alfa-2b trialsHepatology, 1997