The application of electrical impedance tomography to reduce systematic errors in the EEG inverse problem - a simulation study
- 1 August 2000
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 21 (3) , 379-393
- https://doi.org/10.1088/0967-3334/21/3/304
Abstract
In this paper we propose a new method, using the principles of electrical impedance tomography (EIT), to correct for the systematic errors in the inverse problem (IP) of electroencephalography (EEG) that arise from the wrong specification of the electrical conductivities of the head compartments. By injecting known currents into pairs of electrodes and measuring the resulting potential differences recorded from the other electrodes, the equivalent conductivities of brain (σ3), skull (σ2) and scalp (σ1) can be estimated. Since the geometry of the head is assumed to be known, the electrical conductivities remain as the only unknown parameters to be estimated. These conductivities can then be used in the inverse problem of EEG. The simulations performed in this study, using a three-layer sphere to model the head, prove the feasibility of the method, theoretically. Even in the presence of simulated noise with a value of signal-to-noise ratio (SNR) equal to 10, estimations of the electrical conductivities within 5% of the true values were obtained. Simulations showed the existence of a strong relation between errors in the skull thickness and the EIT estimated conductivities. If the skull thickness is wrongly specified, for example overestimated by a factor of two, the conductivity determined by EIT is also overestimated by a factor of two. Simulations showed that this compensation effect also works in the inverse problem of EEG. Application of the proposed method reduces systematic errors in the dipole localization, up to an amount of 1 cm. However it proved to be ineffective to decrease the dipole strength error.Keywords
This publication has 11 references indexed in Scilit:
- The electric resistivity of human tissues (100 Hz-10 MHz): a meta-analysis of review studiesPhysiological Measurement, 1999
- A study of dipole localization accuracy for MEG and EEG using a human skull phantomElectroencephalography and Clinical Neurophysiology, 1998
- Trends in EEG source localizationElectroencephalography and Clinical Neurophysiology, 1998
- The influence of skull-conductivity misspecification on inverse source localization in realistically shaped finite element head modelsBrain Topography, 1997
- Error bounds for EEG and MEG dipole source localizationElectroencephalography and Clinical Neurophysiology, 1993
- Magnetoencephalography—theory, instrumentation, and applications to noninvasive studies of the working human brainReviews of Modern Physics, 1993
- Applied potential tomography: possible clinical applicationsClinical Physics and Physiological Measurement, 1985
- Applied potential tomographyJournal of Physics E: Scientific Instruments, 1984
- EEG Electrode Sensitivity-An Application of ReciprocityIEEE Transactions on Biomedical Engineering, 1969
- The specific resistance of biological material—A compendium of data for the biomedical engineer and physiologistMedical & Biological Engineering & Computing, 1967