Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsis
- 13 January 2004
- journal article
- Published by Springer Nature in Intensive Care Medicine
- Vol. 30 (4) , 709-713
- https://doi.org/10.1007/s00134-003-2116-1
Abstract
To test the effects of low perfusion caused by emerging sepsis on the performance of two new pulse oximetry techniques: Masimo SET in comparison with Nellcor Oxismart XL. Cohort study with random allocation of two pulse oximetry devices to two sensor sites. University animal research facility. Twenty-five adult, anesthetized, ventilated rabbits. Pneumonia/sepsis was induced by tracheal instillation of E. coli. Oxygen saturation was measured by pulse oximetry (SpO2) and recorded continuously until death. Arterial oxygen saturation (SaO2) was measured hourly by CO oximetry and whenever a difference of >5% between the devices occurred. SpO2 sensors were positioned at both forelegs and switched hourly. There was no difference in total signal dropout time [median 3.8 min (range 0.4–66.6 min) vs 3.3 min (range 0–94.5 min), Masimo SET vs Oxismart XL]. There were fewer episodes with a false SpO2 reading [1 (range 0–7) vs 2 (range 0–17)] using the Masimo SET vs the Oxismart XL as verified by CO oximetry; p2–SaO2) was significantly different between the two devices, and variability of bias values increased across time with both devices. Both devices were capable to measure SpO2 during most of the experimental time in this model of low perfusion and therefore appear to be highly sensitive to pick up a signal; however, low perfusion caused by emerging sepsis may result in inaccurate measurements with both devices. These episodes were less common with the Masimo SET vs the Oxismart XL.Keywords
This publication has 15 references indexed in Scilit:
- Pulse oximetry during low perfusion caused by emerging pneumonia and sepsis in rabbitsCritical Care Medicine, 2002
- Reliability of Conventional and New Pulse Oximetry in Neonatal PatientsJournal of Perinatology, 2002
- The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia.2002
- Transesophageal versus surface pulse oximetry in intensive care unit patientsCritical Care Medicine, 2000
- Bacteremia and severe sepsis in adults: a multicenter prospective survey in ICUs and wards of 24 hospitals. French Bacteremia-Sepsis Study Group.American Journal of Respiratory and Critical Care Medicine, 1996
- Signal processing methods for pulse oximetryComputers in Biology and Medicine, 1996
- The accuracy of the biox 3700 pulse oximeter in patients receiving vasoactive therapyIntensive Care Medicine, 1991
- COMPARISON OF FOUR PULSE OXIMETERS: EFFECTS OF VENOUS OCCLUSION AND COLD-INDUCED PERIPHERAL VASOCONSTRICTIONBritish Journal of Anaesthesia, 1990
- Pulse oximetry during hemorrhagic hypotension and cardiopulmonary resuscitation in the rabbitJournal of Critical Care, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986