In Vivo Thrombogenicity of Implantable Cardiac Pacing Leads
- 1 January 1980
- journal article
- research article
- Published by Taylor & Francis in Biomaterials, Medical Devices, and Artificial Organs
- Vol. 8 (4) , 369-379
- https://doi.org/10.3109/10731198009118990
Abstract
Implantable cardiac pacing leads must possess good thromboresistance. Three types of tubular insulator surfaces over the conductor wires were investigated. Materials tested were silicone rubber, smooth and roughened extruded polyurethane. Acute canine intra-arterial implantation methods were used. An index of thrombogenicity based on the weight of total thrombus formed per unit surface area of the vascular implant was determined for each test sample. Scanning electron microscope photographs of the lead surfaces were made. While silicone rubber leads were most thromboresistant, the polyurethane leads performed equally well.Keywords
This publication has 5 references indexed in Scilit:
- Blood-Surface interactions: Fourier transform infrared studies of protein surface adsorption from flowing blood plasma and serumJournal of Biomedical Materials Research, 1979
- Effect of implant surface chemistry upon arterial thrombosisJournal of Biomedical Materials Research, 1979
- Blood Plasma/Implant Interfaces FT—IR Studies of Adsorption on Polyethylene and Heparin‐Treated Polyethylene SurfacesArtificial Organs, 1978
- Applied Chemistry at Protein InterfacesPublished by American Chemical Society (ACS) ,1975
- Development of blood compatible polymers using the electret effectJournal of Biomedical Materials Research, 1971