RF heating of implanted spinal fusion stimulator during magnetic resonance imaging
- 1 May 1997
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 44 (5) , 367-373
- https://doi.org/10.1109/10.568912
Abstract
Radio frequency (RF) heating of an implanted spinal fusion stimulator (SpF) during magnetic resonance imaging (MRI) was studied on a full-size human phantom. Heating during MRI scans (GE Signa 4X, 1.5 T) was measured with RF-transparent fiberoptic sensors. With the implant correctly connected, the maximum temperature rises were less than 2 degrees C during the 26 min that the scans were at maximum RF power. At the tip of a broken stimulator lead (connecting the SpF generator and its electrodes), the maximum temperature rise was 11-14 degrees C. Regular 4-min scans of the spinal cord produced similar temperature rises at the broken tip. After the generator and the leads were removed, heating at the electrode connector tip was less than 1.5 degrees C. The control temperature rises at the same locations, without the stimulator, were less than 0.5 degree C. This study shows that spinal fusion stimulator heating is within the Food and Drug Administration safety guideline of 2 degrees C. However, if a lead wire is broken, it is unsafe during MRI scans. Radiological examinations will be necessary to ensure the integrity of the implant.Keywords
This publication has 11 references indexed in Scilit:
- Ex Vivo Evaluation of Ferromagnetism, Heating, and Artifacts of Breast Tissue Expanders Exposed to a 1.5‐T MR SystemJournal of Magnetic Resonance Imaging, 1995
- Hr imaging and vascular access ports: Ex vivo evaluation of ferromagnetism, heating, and artifacts at 1.5 tJournal of Magnetic Resonance Imaging, 1995
- Absence of radiofrequency heating from auditory implants during magnetic resonance imagingBioelectromagnetics, 1995
- Ex vivo evaluation of ferromagnetism, heating, and artifacts produced by heart valve prostheses exposed to a 1.5‐T MR systemJournal of Magnetic Resonance Imaging, 1994
- Magnetic resonance imaging can cause focal heating in a nonuniform phantomIEEE Transactions on Biomedical Engineering, 1993
- Heating effects of metallic implants by MRI examinationsMagnetic Resonance in Medicine, 1988
- Simulated biological materials for electromagnetic radiation absorption studiesBioelectromagnetics, 1987
- Specific Absorption Rates of Energy in Man Models Exposed to Cellular UHF Mobile-Antenna FieldsIEEE Transactions on Microwave Theory and Techniques, 1986
- Formulas for preparing phantom muscle tissue at various radiofrequenciesBioelectromagnetics, 1984
- Nonionizing electromagnetic wave effects in biological materials and systemsProceedings of the IEEE, 1972