Optimized Sampling in the Gas Chromatography-Infrared Spectroscopy Interface
- 1 July 1977
- journal article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 31 (4) , 284-288
- https://doi.org/10.1366/000370277774463562
Abstract
The optimum volume for a flow-through gas cell used in the gas chromatography-infrared spectroscopy interface (GC-ir) is shown to be equal to the volume of carrier gas between the half-width points of the GC peak. The optimum length and diameter for light-pipe gas cells for GC fractions separated using 1/8 in. packed columns and support-coated open tubular columns has been calculated for infrared measurements made using a rapid-scanning Fourier transform infrared spectrometer with either a triglycine sulfate or a mercury cadmium telluride detector. A sampling trigger actuated by the inflection points of the GC peak has been shown to be superior to the more usual threshold type of trigger. Expressions for the delay time between actuation of the trigger and initiation of data collection are also derived.Keywords
This publication has 4 references indexed in Scilit:
- Rapid Scan Infrared Fourier Transform SpectroscopyApplied Spectroscopy Reviews, 1972
- Infrared Fourier transform spectroscopy in flavor analysis. IV. Spectra of gas chromatography fractionsJournal of Agricultural and Food Chemistry, 1971
- Infrared Examination of Gas Chromatography Effluent using a Dual Beam, Single Detector Interference SpectrometerAnalytical Letters, 1968
- Measurements of infrared spectra of gas-liquid chromatography fractions using multiple-scan interference spectrometryAnalytical Chemistry, 1967