Increased heating by diagnostic ultrasound due to nonlinear propagation
- 30 June 1990
- journal article
- research article
- Published by Acoustical Society of America (ASA) in The Journal of the Acoustical Society of America
- Vol. 88 (1) , 26-34
- https://doi.org/10.1121/1.399950
Abstract
The heating of tissues by the absorption of ultrasound is an important safety consideration in the use of diagnostic ultrasound. This paper shows that models of ultrasonic heating for this situation need to take account of nonlinear propagation. Measurements were made of the temperature rise in a simple of tissue-mimicking gel, caused by the application of 3.6-MHz focused ultrasonic beams for 3 min. The propagation path to the focus was in water, to mimic the situation where the fetus is scanned through the full bladder. The effect of nonlinear propagation was seen by changing the pressure amplitude of the pulse, while altering the pulsing regime to preserve a constant spatial-peak temporal-average intensity of 1 W cm-2. When nonlinear distortion was present, an enhancement in the temperature rise was observed, which correlated with the value of the shock parameter. The enhancement ratio was typically up to a factor of 3, and the maximum temperature rise observed was 2.degree.C. This enhanced heating was seen both at the surface of the tissue-mimicking gel and after propagation through 23 mm of the material. Under conditions of nonlinear propagation, the maximum heating usually occurs in the prefocal region, rather than at the focus.This publication has 11 references indexed in Scilit:
- I n v i t r o ultrasonic heating of fetal boneThe Journal of the Acoustical Society of America, 1989
- Prediction of in situ exposure to ultrasound: An improved methodUltrasound in Medicine & Biology, 1989
- Temperature elevation at muscle-bone interface during scanned, focused ultrasound hyperthermiaInternational Journal of Hyperthermia, 1988
- Demonstration of enhanced temperature elevation due to nonlinear propagation of focussed ultrasound in dog's thigh in vivoUltrasound in Medicine & Biology, 1987
- The development of harmonic distortion in pulsed finite-amplitude ultrasound passing through liverPhysics in Medicine & Biology, 1986
- The output of pulse-echo ultrasound equipment: a survey of powers, pressures and intensitiesThe British Journal of Radiology, 1985
- Thermal conduction errors of manganin-constantan thermocouple arraysPhysics in Medicine & Biology, 1985
- Effects of heat conduction and sample size on ultrasonic absorption measurementsThe Journal of the Acoustical Society of America, 1985
- A theoretical study of nonlinear effects with focused ultrasound in tissues: An “acoustic bragg peak”Ultrasound in Medicine & Biology, 1985
- Evidence for ultrasonic finite-amplitude distortion in muscle using medical equipmentThe Journal of the Acoustical Society of America, 1985