Factors Affecting Accuracy and Precision in PET Volume Imaging
Open Access
- 1 March 1991
- journal article
- Published by SAGE Publications in Journal of Cerebral Blood Flow & Metabolism
- Vol. 11 (2) , A38-A44
- https://doi.org/10.1038/jcbfm.1991.35
Abstract
Volume imaging positron emission tomographic (PET) scanners with no septa and a large axial acceptance angle offer several advantages over multiring PET scanners. A volume imaging scanner combines high sensitivity with fine axial sampling and spatial resolution. The fine axial sampling minimizes the partial volume effect, which affects the measured concentration of an object. Even if the size of an object is large compared to the slice spacing in a multiring scanner, significant variation in the concentration is measured as a function of the axial position of the object. With a volume imaging scanner, it is necessary to use a three-dimensional reconstruction algorithm in order to avoid variations in the axial resolution as a function of the distance from the center of the scanner. In addition, good energy resolution is needed in order to use a high energy threshold to reduce the coincident scattered radiation.Keywords
This publication has 9 references indexed in Scilit:
- Three dimensional reconstruction of PET data from a multi-ring cameraIEEE Transactions on Nuclear Science, 1989
- Cylindrical PET detector designIEEE Transactions on Nuclear Science, 1988
- Simulation studies for cylindrical positron tomographyIEEE Transactions on Nuclear Science, 1988
- The effect of collimation on scatter fraction in multi-slice PETIEEE Transactions on Nuclear Science, 1988
- The Theory of Three-Dimensional Image Reconstruction for PETIEEE Transactions on Medical Imaging, 1987
- An Iterative Image Space Reconstruction Algorthm Suitable for Volume ECTIEEE Transactions on Medical Imaging, 1986
- Spherical positron emission tomograph (S-PET) I - performance analysisNuclear Instruments and Methods in Physics Research, 1984
- Evaluating the Performance of Multiplane Positron Tomographs Designed for Brain ImagingIEEE Transactions on Nuclear Science, 1982
- Design Considerations for Positron Emission TomographyIEEE Transactions on Biomedical Engineering, 1981