We present 32 cases of dysfunction of the Björk–Shiley valvular prosthesis, representing 3.01% of the 1063 BS prostheses implanted at this clinic. The complications observed were thrombosis of the prosthesis, suture dehiscence with or without endocarditis and discsuture interference. In selected, favourable cases, we prefer thrombectomy with generous washing of the prosthesis and rotation of the disc, without removal of the disc from the prosthesis. Furthermore, when the dehiscence does not exceed 1/3 of the circumference of the ring, we have achieved very good results with single sutures supported on Teflon patches. Subsequent failures of this procedure were not of a technique nature, but rather due to the poor condition of the patients. In situations of clinical emergency, the diagnosis of valvular dysfunction may suffice as indication of surgery, further diagnostic measures being postponed in view of the data of the physical exploration.This publication has 26 references indexed in Scilit:
- Thrombosis of the Björk-Shiley Aortic Valve ProsthesisPublished by American Medical Association (AMA) ,1978
- Arterial thromboembolic complications in patients with Björk-Shiley and Lillehei-Kaster aortic disc valve prosthesesAmerican Heart Journal, 1977
- Reoperation for dysfunction of the Björk-Shiley mitral disc prosthesisAmerican Heart Journal, 1977
- Malfunction of Bjork-Shiley valve prosthesis in tricuspid position.Heart, 1976
- Late Thrombosis of the Aortic Bjork-Shiley ProsthesisChest, 1976
- Echocardiographic Diagnosis of a Stuck Bjork-Shiley Aortic Valve ProsthesisChest, 1976
- The Björk-Shiley Mitral Valve Prosthesis: A Comparative Study with Different Prosthesis OrientationsThe Annals of Thoracic Surgery, 1974
- Results of Use of the Pyrolytic Carbon Tilting Disc Bjork-Shiley Aortic ProsthesisChest, 1974
- Encapsulation of the Björk-Shiley Aortic Disc Valve Prosthesis Caused by the Lack of Anticoagulation TreatmentScandinavian Journal of Thoracic and Cardiovascular Surgery, 1973
- Disc Wear in the Björk-Shiley Tilting Disc ValveScandinavian Journal of Thoracic and Cardiovascular Surgery, 1971