Doppler echocardiographic characteristics of normal and dysfunctioning prosthetic valves in the tricuspid and mitral position.
Open Access
- 1 January 1990
- Vol. 63 (1) , 41-44
- https://doi.org/10.1136/hrt.63.1.41
Abstract
The Doppler echocardiographic characteristics of 70 prosthetic valves in 35 patients are reported. Twenty nine patients had a Björk-Shiley prosthesis in both mitral and tricuspid positions and six had Carpentier-Edwards valves in both sites. Five of the patients had abnormal tricuspid prostheses on the basis of clinical and echocardiographic criteria. Pulsed wave Doppler echocardiography was used in all examinations. The pressure half times for the normal tricuspid prosthetic valves, 105 (40) ms (Björk-Shiley) and 97 (53) ms (Carpentier-Edwards), were significantly longer than those of normal mitral prosthetic valves, 75 (18) ms (Björk-Shiley) and 83 (15) ms (Carpentier-Edwards). The range of pressure half times for the abnormal tricuspid valves (237-530 ms) was distinct from that of the apparently normal tricuspid prosthetic valves (38-197 ms). There was an increase in the peak velocity of the obstructed tricuspid prosthetic valves (1.69 (0.12) m/s) in comparison with normal prostheses (1.06 (0.26) m/s). The normal range of pressure half times for the Björk-Shiley and Carpentier-Edwards valves in the mitral position is not applicable to the same valves in the tricuspid position. The valve appears to function well with very long pressure half times but a pressure half time of greater than 200 ms coupled with a peak velocity of greater than 1.60 ms without significant valve regurgitation indicates tricuspid obstruction of the tricuspid prosthetic valve.This publication has 12 references indexed in Scilit:
- Doppler Echocardiography for Assessing Left Ventricular Diastolic FunctionAnnals of Internal Medicine, 1988
- Doppler and echocardiographic features of normal and dysfunctioning bioprosthetic valvesJournal of the American College of Cardiology, 1987
- Right ventricular filling detected by pulsed Doppler echocardiography during the convalescent stage of inferior wall acute myocardial infarctionThe American Journal of Cardiology, 1987
- Validation of continuous-wave Doppler echocardiographic measurements of mitral and tricuspid prosthetic valve gradients: a simultaneous Doppler-catheter study.Circulation, 1986
- Normal and abnormal prosthetic valve function as assessed by doppler echocardiographyJournal of the American College of Cardiology, 1986
- Doppler echocardiographic characteristics of the Carpentier-Edwards xenograftEuropean Heart Journal, 1986
- Doppler hemodynamic evaluation of prosthetic (Starr-Edwards and Björk-Shiley) and Bioprosthetic (Hancock and Carpentier-Edwards) cardiac valvesThe American Journal of Cardiology, 1985
- Normal intracardiac and great artery blood velocity measurements by pulsed Doppler echocardiography.Heart, 1985
- Baseline rest electrocardiographic abnormalities, antihypertensive treatment, and mortality in the Multiple Risk Factor Intervention TrialThe American Journal of Cardiology, 1985
- Noninvasive assessment of atrioventricular pressure half-time by Doppler ultrasound.Circulation, 1979