Design of a visible-light spectroscopy clinical tissue oximeter

Abstract
We develop a clinical visible-light spectroscopy (VLS) tissue oximeter. Unlike currently approved near-infrared spectroscopy (NIRS) or pulse oximetry (SpO2%), VLS relies on locally absorbed, shallow-penetrating visible light (475 to 625 nm) for the monitoring of microvascular hemoglobin oxygen saturation (StO2%), allowing incorporation into therapeutic catheters and probes. A range of probes is developed, including noncontact wands, invasive catheters, and penetrating needles with injection ports. Data are collected from: 1. probes, standards, and reference solutions to optimize each component; 2. ex vivo hemoglobin solutions analyzed for StO2% and pO2 during deoxygenation; and 3. human subject skin and mucosal tissue surfaces. Results show that differential VLS allows extraction of features and minimization of scattering effects, in vitro VLS oximetry reproduces the expected sigmoid hemoglobin binding curve, and in vivo VLS spectroscopy of human tissue allows for real-time monitoring (e.g., gastrointestinal mucosal saturation 69±4%, n=804; gastrointestinal tumor saturation 45±23%, n=14; and p