Initial Experience with a New Single Chamber, Dual Sensor Rate Responsive Pacemaker
- 1 September 1993
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 16 (9) , 1833-1841
- https://doi.org/10.1111/j.1540-8159.1993.tb01818.x
Abstract
In August 1991, a new single chamber pacemaker became available that utilizes information from two sensors, activity and stimulus-to-T wave (QT) interval. We are reporting on the first 90 implants in 21 centers. T wave sensing was adequate at implantation in 88/90 patients, with a safety margin of > 100% in 86/90. Activity sensing was adequate in all patients. The contribution of each sensor (sensor blending) is programmable for each patient. Of 75 patients assessed at 1 month after implant, three have been programmed to ''Activity-Only'' mode, and 72 to dual sensor mode. Of these, 18 have been programmed to ''QT < Activity,'' 48 to ''QT = Activity,'' and 6 to ''QT > Activity.'' Forty-five patients underwent exercise testing in dual sensor mode and a subgroup of 15 also underwent exercise testing in Activity-Only mode. The dual sensor mode produced a more gradual increase in pacing rate. Sensor Cross Checking(TM) satisfactorily prevented a sustained high pacing rate in tests of false-positive activity sensing (tapping, vibrating pacemaker, or static pressure). The maximum pacing rate on walking downstairs (94.2 +/- 7.2 ppm) was similar to that produced by walking upstairs (91.6 +/- 5.9 ppm). We conclude that initial assessment of this dual sensor, single chamber, rate responsive pacemaker confirms that the algorithm for combining data from two sensors functions satisfactorily. Dual sensor rate responsive pacing may offer significant advantages over single sensor devices, and further studies of this novel device are indicatedKeywords
This publication has 25 references indexed in Scilit:
- Standardized Informal Exercise Testing for Programming Rate Adaptive PacemakersPacing and Clinical Electrophysiology, 1991
- Applicability of the Stimulus‐T Interval for Antitachycardia PacingPacing and Clinical Electrophysiology, 1991
- One‐Year Follow‐Up of Automatic Adaptation of the Rate Response Algorithm of the QT Sensing, Rate Adaptive PacemakerPacing and Clinical Electrophysiology, 1991
- Initial Clinical Experience with Rate Adaptive Cardiac Pacing Using Two Sensors SimultaneouslyPacing and Clinical Electrophysiology, 1990
- Rate-modulated pacing.Circulation, 1990
- QT Sensing Rate Responsive Pacing Compared to Fixed Rate Ventricular Inhibited Pacing: A Controlled Clinical StudyPacing and Clinical Electrophysiology, 1989
- A New Rate-Modulated Pacemaker System Optimized by Combination of Two SensorsPacing and Clinical Electrophysiology, 1988
- Single-chamber cardiac pacing with activity-initiated chronotropic response: evaluation by cardiopulmonary exercise testing.Circulation, 1987
- Activity‐Sensing, Rate‐Responsive Pacing: Improvement in Myocardial Performance with ExercisePacing and Clinical Electrophysiology, 1985
- The Use of QT Interval to Determine Pacing Rate: Early Clinical ExperiencePacing and Clinical Electrophysiology, 1983