Removal of AC interference from the electrocardiogram.
Open Access
- 31 July 1977
- Vol. 39 (8) , 907-908
- https://doi.org/10.1136/hrt.39.8.907
Abstract
A method is described for the removal of AC interference from the electrocardiogram in monitors which do not have isolated inputs. The patient's leads are connected to an isolation amplifier which provides isolation between the input and the output and has a high common mode rejection ratio. The output from the isolation amplifier is taken to the electrocardiographic monitor. The isolation amplifier not only provides additional patient safety but also reduces the level of AC interference without interfering with any frequency components in the electrocardiogram.Keywords
This publication has 2 references indexed in Scilit:
- Disturbances of heart rhythm produced by 50 Hz leakage currents in dogsCardiovascular Research, 1975
- Safe use of electro–medical equipmentAnaesthesia, 1971