Safety Profiles and Lesion Size of Different Radiofrequency Ablation Technologies: A Comparison of Large Tip, Open and Closed Irrigation Catheters
- 20 February 2009
- journal article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 20 (3) , 325-335
- https://doi.org/10.1111/j.1540-8167.2008.01305.x
Abstract
Different technologies have been developed for radiofrequency ablation (RFA), which include increasing electrode (tip) size and cooling the tip through irrigation either internally (closed-loop) with D5W or externally (open-loop) with saline. Although these catheters are widely used clinically, the propensity for adverse events and the lesion profiles of each of these catheter technologies have not been directly compared under a wide range of controlled conditions. Freshly excised canine thigh muscle was placed in a chamber filled with circulating, heparinized blood heated to 37 degrees C. Five different catheters were tested: 4 mm tip, 10 mm tip single thermistor, 10 mm tip multitemperature sensor, 4 mm closed-loop irrigated cooled-tip, and 4 mm open-loop irrigated cooled tip at several different contact and power settings. The catheter and tissue interface was continuously monitored with intracardiac echocardiography (echo) (Acuson). During the RFA, any bubbling generated from the tip and/or popping seen on echo was noted, and after each RFA, the catheter and lesion were examined for the presence of thrombus. For all of the catheters, complications correlated to the electrode tip temperature and power setting. All of the catheters experienced complications at any lesion size except for the open-irrigated catheter, which only had complications at the largest lesions. Overall, the cooled tip catheters experienced an at least sixfold greater odds of popping, bubbling, and impedance rises than the 4 mm, but the majority occurred at power levels greater than 20 W. The open-irrigated catheters created eccentric lesions that extended away from the tissue-catheter interface, in the direction of blood flow. In addition, it produced saline filled blisters at the lesion site in 16.7% of the burns. The 10 mm catheter had an at least twofold greater odds of thrombus, charring, and bubbling, but larger lesions than the 10 mm multitemperature sensor catheter. Catheter type, contact conditions, and power settings all play a role in lesion size and in the frequency of complications that occur during an RFA. Cooling the electrode tip, either internally or externally, does not prevent complications from occurring, especially at the higher power control settings. Adding more temperature sensors to the 10 mm seems to reduce the amount of complications that can occur.Keywords
This publication has 25 references indexed in Scilit:
- Electrophysiologic and Anatomic Characterization of Sites Resistant to Electrical Isolation During Circumferential Pulmonary Vein Ablation for Atrial Fibrillation: A Prospective StudyJournal of Cardiovascular Electrophysiology, 2007
- Impedance and Temperature Monitoring Improve the Safety of Closed‐Loop Irrigated‐Tip Radiofrequency AblationJournal of Cardiovascular Electrophysiology, 2007
- Incidence, Time Course, and Characteristics of Microbubble Formation During Radiofrequency Ablation of Pulmonary Veins with an 8‐mm Ablation CatheterPacing and Clinical Electrophysiology, 2006
- Avoiding Microbubbles Formation During Radiofrequency Left Atrial Ablation Versus Continuous Microbubbles Formation and Standard Radiofrequency Ablation Protocols: Comparison of Energy Profiles and Chronic Lesion CharacteristicsJournal of Cardiovascular Electrophysiology, 2006
- Microbubbles during radiofrequency catheter ablation: Composition and formationHeart Rhythm, 2005
- Optimizing RF Output for Cooled RF AblationJournal of Cardiovascular Electrophysiology, 2004
- Ablation with an internally irrigated radiofrequency catheter: Learning how to avoid steam popsHeart Rhythm, 2004
- Mechanisms for Enlarging Lesion Size During Irrigated Tip Radiofrequency Ablation:. Is There a Virtual Electrode Effect?Journal of Interventional Cardiology, 2004
- Radiofrequency Delivery Through a Cooled Catheter Tip Allows the Creation of Larger Endomyocardial Lesions in the Ovine HeartJournal of Cardiovascular Electrophysiology, 1995
- Use of the Saline Infusion Electrode Catheter for Improved Energy Delivery and Increased Lesion Size in Radiofrequency Catheter AblationPacing and Clinical Electrophysiology, 1995