Effect of central corneal thickness on Goldmann applanation tonometry measures—a different result with different pachymeters
- 24 May 2007
- journal article
- research article
- Published by Springer Nature in Albrecht von Graefes Archiv für Ophthalmologie
- Vol. 245 (11) , 1603-1610
- https://doi.org/10.1007/s00417-007-0601-x
Abstract
To assess whether the use of Orbscan pachymetry, as an alternative to ultrasound pachymetry, had an effect on the IOP-CCT relationship for Goldmann tonometry in normal human eyes. The right eye of 50 healthy subjects, aged 19 to 82 years, were assessed by Orbscan (three scans), and then central corneal thickness (CCT) measured by ultrasound pachymetry (three readings) and Goldmann tonometry performed (three readings). Goldmann tonometry indicated average IOP values from 9.7 to 20 mm Hg (group mean 14.3 ± 2.5 mm Hg) that were highly highly correlated with ultrasound pachymetry (r = 0.608, p < 0.001). Orbscan central zone pachymetry (2-mm sample zone) yielded higher values than ultrasound, averaging 0.591 ± 0.044 mm compared to 0.523 ± 0.037 mm (p < 0.001), with the differences between the instruments being proportional to the average thickness (p < 0.001; r = 0.461). Notwithstanding, the Goldmann tonometry values were still highly correlated with the Orbscan central zone thickness data (r = 0.595, p < 0.001); but, due to the difference in the thickness data generated by the two pachymeters, the absolute slope of the IOP–CCT relationship was slightly less if Orbscan measures were used, with or without use of the acoustic factor. Orbscan measures of the thickness of a central corneal zone can be used to assess the impact of central corneal thickness on Goldmann tonometry data.Keywords
This publication has 21 references indexed in Scilit:
- On the use of Orbscan II to assess the peripheral corneal thickness in humans: a comparison with ultrasound pachometry measuresOphthalmic and Physiological Optics, 2007
- The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometryBritish Journal of Ophthalmology, 2005
- Central and paracentral corneal pachymetry in patients with normal tension glaucoma and ocular hypertensionAlbrecht von Graefes Archiv für Ophthalmologie, 2005
- Orbscan computerized topography: Attributes, applications, and limitationsJournal of Cataract & Refractive Surgery, 2005
- Corneal pachymetry in normal and keratoconic eyesJournal of Cataract & Refractive Surgery, 2004
- Central corneal thickness in European (white) individuals, especially children and the elderly, and assessment of its possible importance in clinical measures of intra‐ocular pressureOphthalmic and Physiological Optics, 2002
- Human Corneal Thickness and Its Impact on Intraocular Pressure MeasuresSurvey of Ophthalmology, 2000
- Corneal Indentation During Ultrasonic PachometryCornea, 1999
- Sources of error with use of Goldmann-type tonometersSurvey of Ophthalmology, 1993
- Über ApplanationstonometrieOphthalmologica, 1957