Corneal pachymetry in normal and keratoconic eyes
- 1 June 2004
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 30 (6) , 1272-1277
- https://doi.org/10.1016/j.jcrs.2003.11.049
Abstract
Purpose: To compare corneal thickness measurements using Orbscan II (OII) and ultrasonic (US) pachymetry in normal and in keratoconic eyes. Setting: Eye Department, Heartlands and Solihull NHS Trust, Birmingham, United Kingdom. Methods: Central corneal thickness (CCT) was measured by means of OII and US pachymetry in 1 eye of 72 normal subjects and 36 keratoconus patients. The apical corneal thickness (ACT) in keratoconus patients was also evaluated using each method. The mean of the difference, standard deviation (SD), and 95% limits of agreement (LoA = mean ± 2 SD), with and without applying the default linear correction factor (LCF), were determined for each sample. The Student t test was used to identify significant differences between methods, and the correlation between methods was determined using the Pearson bivariate correlation. Bland-Altman analysis was performed to confirm that the results of the 2 instruments were clinically comparable. Results: In normal eyes, the mean difference (± 95% LoA) in CCT was 1.04 μm ± 68.52 (SD) (P>.05; r = 0.71) when the LCF was used and 46.73 ± 75.40 μm (P = .0001; r = 0.71) without the LCF. In keratoconus patients, the mean difference (± 95% LoA) in CCT between methods was 42.46 ± 66.56 μm (P<.0001: r = 0.85) with the LCF, and 2.51 ± 73.00 μm (P>.05: r = 0.85) without the LCF. The mean difference (± 95% LoA) in ACT for this group was 49.24 ± 60.88 μm (P<.0001: r = 0.89) with the LCF and 12.71 ± 68.14 μm (P = .0077; r = 0.89) when the LCF was not used. Conclusions: This study suggests that OII and US pachymetry provide similar readings for CCT in normal subjects when an LCF is used. In keratoconus patients, OII provides a valid clinical tool for the noninvasive assessment of CCT when the LCF is not applied.Keywords
This publication has 17 references indexed in Scilit:
- Comparison of corneal pachymetry using ultrasound and Orbscan IIJournal of Cataract & Refractive Surgery, 2002
- Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyesJournal of Cataract & Refractive Surgery, 2001
- Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement proceduresJournal of Cataract & Refractive Surgery, 2001
- Keratoconus evaluation using the Orbscan Topography SystemJournal of Cataract & Refractive Surgery, 2000
- Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography systemBritish Journal of Ophthalmology, 1999
- Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetryJournal of Cataract & Refractive Surgery, 1997
- Micrometer-Scale Resolution Imaging of the Anterior Eye In Vivo With Optical Coherence TomographyArchives of Ophthalmology (1950), 1994
- Interferometric Measurement of Corneal Thickness With Micrometer PrecisionAmerican Journal of Ophthalmology, 1994
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- THE THICKNESS OF THE HUMAN CORNEA AS DETERMINED BY A SPECULAR METHODActa Ophthalmologica, 1984