Abstract
We have measured the local blood flow (BF) and oxygen consumption (OC) in the human calf muscle using near-infrared spectroscopy during venous occlusion. Venous occlusion was achieved by inflating a pneumatic cuff around the thigh of the subject. We have investigated the influence of the inflation time and cuff pressure on the recovered values of BF and OC. We have found that if the cuff pressure is increased from a threshold pressure (approximately 30 mm Hg) to a critical pressure (approximately 45 mm Hg) in less than about 6 s, one measures the same values of BF and OC independent of the total inflation time and final cuff pressure. We also report nine-pixel spatial maps of BF and OC to show that this technique can lead to spatially resolved measurements of blood flow and oxygen consumption in tissues.Keywords
This publication has 19 references indexed in Scilit:
- Near-infrared spectroscopy and the tilting table protocol: a novel method to study the blood flow and the oxygen consumption in tissuesPublished by SPIE-Intl Soc Optical Eng ,1999
- Measurement of Cerebral Oxygen Consumption in the Human Neonate Using Near Infrared Spectroscopy: Cerebral Oxygen Consumption Increases with Advancing Gestational AgePediatric Research, 1998
- Validation of measurement protocols to assess oxygen consumption and blood flow in the human forearm by near-infrared spectroscopyPublished by SPIE-Intl Soc Optical Eng ,1998
- Comparison of two methods of measuring forearm oxygen consumption (VO2) by near infrared spectroscopyJournal of Biomedical Optics, 1997
- Measurement of Cerebral Venous Oxyhemoglobin Saturation in Children by Near-Infrared Spectroscopy and Partial Jugular Venous OcclusionPediatric Research, 1995
- Noninvasive measurement of forearm blood flow and oxygen consumption by near-infrared spectroscopyJournal of Applied Physiology, 1994
- Measurement of hemoglobin flow and blood flow by near-infrared spectroscopyJournal of Applied Physiology, 1993
- Estimation of Cerebral Venous Saturation in Newborn Infants by Near Infrared SpectroscopyPediatric Research, 1993
- Superposition of Arteriolar Vasomotion Waves and Regulation of Blood Flow in Skeletal Muscle MicrocirculationPublished by Springer Nature ,1990
- Oxygen Transport to Tissue during Recurrent Blood Flow Supply by Grouped Capillaries in Skeletal Muscle with or without Facilitated DiffusionPublished by Springer Nature ,1988