An automated system for testing the accuracy of patient‐controlled analgesia devices
- 1 August 1992
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 47 (8) , 693-696
- https://doi.org/10.1111/j.1365-2044.1992.tb02394.x
Abstract
A system was developed to test the accuracy of patient-controlled analgesia devices in situations simulating clinical use. Bolus requests are made automatically at predetermined intervals, and the infusate delivered is measured and recorded without the need for operator presence. To ensure clinical relevance, the bolus request times used in this study corresponded to a pattern typical of those requested by patients on the ward. Graseby, Abbott Provider 5500 and IVAC patient-controlled analgesia devices were tested and found to deliver reasonably accurately over a 24 h period. However, when an infusion was started in an unprimed system or after a period of no bolus requests in a bolus-only mode the Graseby and IVAC machines under-delivered. This system provides a means of testing patient-controlled analgesia devices operating in any delivery mode.Keywords
This publication has 6 references indexed in Scilit:
- Evaluation of the Graseby PCASAnaesthesia, 1991
- Assessment of a New Infusion Pump for Epidural PCAAnaesthesia and Intensive Care, 1991
- Patient-controlled analgesia — does it provide more than comfort?Canadian Journal of Anesthesia/Journal canadien d'anesthésie, 1990
- Patient-Controlled Analgesia: A Review of Effectiveness of Therapy and an Evaluation of Currently Available DevicesDICP, 1989
- Patient‐controlled analgesiaAnaesthesia, 1986
- Objective Measurement of PainAnesthesiology, 1968