Measurement of acoustic output parameters from medical ultrasound devices with an RF power meter system
- 1 March 1988
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 35 (2) , 140-145
- https://doi.org/10.1109/58.4163
Abstract
Current national and international standards for the measurement of acoustic output parameters of diagnostic medical ultrasound devices rely for their strict implementation on the ability to measure pulse intensity integral or pressure squared integral. A relatively inexpensive measurement system that utilizes a calibrated hydrophone in conjunction with a thermal RMS device is described. It is shown that this system can be used to locate maximum pulse average parameters in static beam devices and true maximum temporal average parameters in complex real-time diagnostic systems. The advantages of the system are discussed and measurement uncertainties are assessed.<>Keywords
This publication has 5 references indexed in Scilit:
- Measurement and characterisation of the acoustic output of medical ultrasonic equipment Part 2Medical & Biological Engineering & Computing, 1986
- The locations of peak pressures and peak intensities in finite amplitude beams from a pulsed focused transducerUltrasound in Medicine & Biology, 1986
- A discussion of procedures for ultrasonic intensity and power calculations from miniature hydrophone measurementsUltrasound in Medicine & Biology, 1985
- Acoustic shock generation by ultrasonic imaging equipmentThe British Journal of Radiology, 1984
- PVDF membrane hydrophone performance properties and their relevance to the measurement of the acoustic output of medical ultrasonic equipmentJournal of Physics E: Scientific Instruments, 1983