Cerebral embolism has been implicated in the development of cognitive and neurological deficits following bypass surgery. This study proposes methodology for estimating cerebral air embolus volume using transcranial Doppler sonography. Transcranial Doppler audio signals of air bubbles in the middle cerebral artery obtained from in vivo experiments were subjected to a fast-Fourier transform analysis. Audio segments when no air was present as well as artifact resulting from electrocautery and sensor movement were also subjected to fast-Fourier transform analysis. Spectra were compared, and frequency and power differences were noted and used for development of audio band-pass filters for isolation of frequencies associated with air emboli. In a bench model of the middle cerebral artery circulation, repetitive injections of various air volumes between 0.5 and 500 microL were made. Transcranial Doppler audio output was band-pass filtered, acquired digitally, then subjected to a fast-Fourier transform power spectrum analysis and power spectrum integration. A linear least-squares correlation was performed on the data. Fast-Fourier transform analysis of audio segments indicated that frequencies between 250 and 500 Hz are consistently dominant in the spectrum when air emboli are present. Background frequencies appear to be below 240 Hz, and artifact resulting from sensor movement and electrocautery appears to be below 300 Hz. Data from the middle cerebral artery model filtered through a 307- to 450-Hz band-pass filter yielded a linear relation between emboli volume and the integrated value of the power spectrum near 40 microL. Detection of emboli less than 0.5 microL was inconsistent, and embolus volumes greater than 40 microL were indistinguishable from one another. The preliminary technique described in this study may represent a starting point from which automated detection and volume estimation of cerebral emboli might be approached.Keywords
This publication has 23 references indexed in Scilit:
- NEUROPSYCHOLOGICAL CHANGE AND STATISTICAL ARTIFACTJournal of Neurosurgical Anesthesiology, 1991
- Clinical comparison of two devices for detection of microemboli during cardiopulmonary bypassClinical Physics and Physiological Measurement, 1990
- Doppler ultrasound estimation of microbubbles in the arterial line during extracorporeal circulationPerfusion, 1990
- The Transcranial Doppler Can Image Microaggregates of Intracranial Air and Particulate MatterJournal of Neurosurgical Anesthesiology, 1989
- The Detection of Microemboli in the Middle Cerebral Artery during Cardiopulmonary Bypass: A Transcranial Doppler Ultrasound Investigation Using Membrane and Bubble OxygenatorsThe Annals of Thoracic Surgery, 1987
- Comparison of Bubble Release from Various Types of OxygenatorsScandinavian Journal of Thoracic and Cardiovascular Surgery, 1985
- Gas Bubble Detection in Fluid Lines by Means of Pulsed Doppler UltrasoundScandinavian Journal of Thoracic and Cardiovascular Surgery, 1985
- Doppler Ultrasound Estimation of Bubble Removal by Various Arterial Line Filters During Extracorporeal CirculationScandinavian Journal of Thoracic and Cardiovascular Surgery, 1982
- The Use of Ultrasonics in the Determination of Arterial Aeroembolism During Open-Heart SurgeryThe Annals of Thoracic Surgery, 1969