An evaluation of the Vita-Stat automatic blood pressure measuring device.
- 1 March 1980
- journal article
- abstracts
- Published by Wolters Kluwer Health in Hypertension
- Vol. 2 (2) , 221-227
- https://doi.org/10.1161/01.hyp.2.2.221
Abstract
Four Vita-Stat (VS) automatic, coin-operated, blood pressure measuring devices were evaluated for accuracy and precision. Under field conditions, 342 adults had two blood pressure measurements with both the VS device and with the Random-Zero (RZ) device. Two of the VS machines gave significantly higher systolic (SBP) and diastolic blood pressure (DBP) values, compared to the RZ values. The mean difference between these two units and the RZ device were clinically important (14.0 and 6.9 mm Hg SBP; 7.5 and 6.8 mm Hg DBP), and resulted in the misclassification of 23% of normotensives as hypertensives. We observed a significant order effect on SBP with the VS device, with a mean decrement of 4.9 mm Hg between the first and second SBP values, compared to 0.6 mm Hg with the RZ device. Even after adjusting for this bias, all four VS devices gave significantly more variable SBP readings than the RZ unit; two of the four also gave more variable DBP values. These data suggest that the VS device is neither accurate nor precise enough at the present time to be recommended for widespread use. These findings also raise questions concerning the monitoring of performance of this and similar devices in the field.Keywords
This publication has 5 references indexed in Scilit:
- Evaluation of an automated blood pressure measuring device intended for general public use.American Journal of Public Health, 1979
- Variability of blood pressure and the results of screening in the Hypertension Detection and Follow-up ProgramJournal of Chronic Diseases, 1978
- Increased Absenteeism from Work after Detection and Labeling of Hypertensive PatientsNew England Journal of Medicine, 1978
- Evaluation of performance of selected devices for measuring blood pressureThe American Journal of Cardiology, 1973
- STANDARDISATION OF OBSERVERS IN BLOOD-PRESSURE MEASUREMENTThe Lancet, 1965