The interobserver reliability and validity of volume calculation from three‐dimensional ultrasound datasets in the in vitro setting
Top Cited Papers
- 14 March 2003
- journal article
- research article
- Published by Wiley in Ultrasound in Obstetrics & Gynecology
- Vol. 21 (3) , 283-291
- https://doi.org/10.1002/uog.61
Abstract
Objectives The primary aim of this validation study was to determine the interobserver reliability and validity of measurements of phantom objects of known volume using conventional and rotational techniques of volume calculation according to measurement technique. Methods Two observers each acquired a single three‐dimensional ultrasound dataset of three water‐filled objects of different size and shape. The same two observers measured all six datasets using both the conventional technique and the newer rotational technique (Virtual Organ Computer‐aided AnaLysis, VOCAL™) of volume calculation. Reliability was assessed by calculating intraclass correlation coefficients (ICC) and validity by examining the percentage difference from the ‘true’ volume, as determined by a water displacement technique, by the limits of agreement method. Results All of the techniques were highly reliable (ICC: 0.9962–0.9997) and valid to within 4% of the ‘true’ volumes. There were no significant differences in reliability according to measurement plane or between observers. Measurements made with the 6° rotation step were significantly more reliable than those made by all other techniques with the exception of the 9° rotation step (P < 0.05) and significantly more valid than those made with the 30° rotation step or conventional technique (P < 0.05). Conclusions Volume calculation in the in vitro setting is both reliable and valid but is dependent upon the technique applied, with rotational measurements of volume proving superior to conventional techniques. Copyright © 2003 ISUOG. Published by John Wiley & Sons, Ltd.Keywords
Funding Information
- Diagnostic Sonar Limited
This publication has 24 references indexed in Scilit:
- The reproducibility of endometrial volume acquisition and measurement with the VOCAL‐imaging programUltrasound in Obstetrics & Gynecology, 2002
- Reproducibility of transvaginal three‐dimensional endometrial volume measurements with virtual organ computer‐aided analysis (VOCAL) during ovarian stimulationUltrasound in Obstetrics & Gynecology, 2002
- Evaluation of a clinical test. I: Assessment of reliabilityBritish Journal of Obstetrics and Gynaecology, 2001
- The reliability and validity of three dimensional ultrasound volumetric measurements using an in vitro balloon and in vivo uterine modelBJOG: An International Journal of Obstetrics and Gynaecology, 2001
- Evaluation of a clinical test. II: Assessment of validityBJOG: An International Journal of Obstetrics and Gynaecology, 2001
- Ovarian Volume Related to AgeGynecologic Oncology, 2000
- Accuracy of Three-Dimensional Ultrasonography in Volume Estimation of Cervical CarcinomaGynecologic Oncology, 1997
- Oedema formation with the vasodilators nifedipine and diazoxideJournal Of Hypertension, 1996
- Three‐dimensional ultrasound: accuracy of distance and volume measurementsUltrasound in Obstetrics & Gynecology, 1996
- Forming inferences about some intraclass correlation coefficients.Psychological Methods, 1996