Treatment head design for multileaf collimated high-energy electrons
- 1 October 1999
- journal article
- Published by Wiley in Medical Physics
- Vol. 26 (10) , 2161-2167
- https://doi.org/10.1118/1.598732
Abstract
This paper describes how a conventional treatment head can be modified for use of multileaf collimated electron beams. Automatic and dynamic beam delivery are possible for both electrons and photons by using the computer controlled multileaf collimator (MLC) for both photon and electron beams. Thereby, the electron beams can be mixed more freely into the treatment to take advantage of the specific depth modulation characteristics of electrons. The investigation was based on Monte Carlo calculations using the software package BEAM. The physical parameters used in this optimization were the beam penumbra and the virtual/effective point source position. These parameters are essential for shaping beams, beam matching and for dosimetry calculations. The optimization was carried out by modifying a number of parameters: replacing the air atmosphere in the treatment head with helium, adding a helium bag below the MLC, changing the position of the scattering foils, modifying the monitor chamber, and adjusting the position of the MLC. The beam characteristics for some of these designs were found to fulfil our criteria for clinically useful beams down to at least 9 MeV.Keywords
This publication has 14 references indexed in Scilit:
- Methods for the determination of effective monitor chamber thicknessMedical Physics, 1999
- Characteristics of scattered electron beams shaped with a multileaf collimatorMedical Physics, 1997
- Feasibility study of multileaf collimated electrons with a scattering foil based acceleratorRadiotherapy and Oncology, 1996
- Matching of electron beams for conformal therapy of target volumes at moderate depthsRadiotherapy and Oncology, 1996
- Electron mass scattering powers: Monte Carlo and analytical calculationsMedical Physics, 1995
- BEAM: A Monte Carlo code to simulate radiotherapy treatment unitsMedical Physics, 1995
- Matching of electron and photon beams with a multi-leaf collimatorRadiotherapy and Oncology, 1993
- Clinical electron‐beam dosimetry: Report of AAPM Radiation Therapy Committee Task Group No. 25Medical Physics, 1991
- Design Principles and Clinical Possibilities with A New Generation of Radiation Therapy Equipment: A ReviewActa Oncologica, 1987
- Presta: The parameter reduced electron-step transport algorithm for electron monte carlo transportNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1986