A novel four-dimensional radiotherapy method for lung cancer: imaging, treatment planning and delivery

Abstract
We present treatment planning methods based on four-dimensional computed tomography (4D-CT) to incorporate tumour motion using (1) a static field and (2) a dynamic field. Static 4D fields are determined to include the target in all breathing phases, whereas dynamic 4D fields are determined to follow the shape of the tumour assessed from 4D-CT images with a dynamic weighting factor. The weighting factor selection depends on the reliability of patient breathing and limitations of the delivery system. The static 4D method is compared with our standard protocol for gross tumour volume (GTV) coverage, mean lung dose and V20. It was found that the GTV delineated on helical CT without incorporating breathing motion does not adequately represent the target compared to the GTV delineated from 4D-CT. Dosimetric analysis indicates that the static 4D-CT based technique results in a reduction of the mean lung dose compared with the standard protocol. Measurements on a moving phantom and simulations indicated that 4D radiotherapy (4D-RT) synchronized with respiration-induced motion further reduces mean lung dose and V20, and may allow safe application of dose escalation and CRT/IMRT. The motions of the chest cavity, tumour and thoracic structures of 24 lung cancer patients are also analysed.