A novel four-dimensional radiotherapy method for lung cancer: imaging, treatment planning and delivery
- 6 June 2006
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (12) , 3251-3267
- https://doi.org/10.1088/0031-9155/51/12/017
Abstract
We present treatment planning methods based on four-dimensional computed tomography (4D-CT) to incorporate tumour motion using (1) a static field and (2) a dynamic field. Static 4D fields are determined to include the target in all breathing phases, whereas dynamic 4D fields are determined to follow the shape of the tumour assessed from 4D-CT images with a dynamic weighting factor. The weighting factor selection depends on the reliability of patient breathing and limitations of the delivery system. The static 4D method is compared with our standard protocol for gross tumour volume (GTV) coverage, mean lung dose and V20. It was found that the GTV delineated on helical CT without incorporating breathing motion does not adequately represent the target compared to the GTV delineated from 4D-CT. Dosimetric analysis indicates that the static 4D-CT based technique results in a reduction of the mean lung dose compared with the standard protocol. Measurements on a moving phantom and simulations indicated that 4D radiotherapy (4D-RT) synchronized with respiration-induced motion further reduces mean lung dose and V20, and may allow safe application of dose escalation and CRT/IMRT. The motions of the chest cavity, tumour and thoracic structures of 24 lung cancer patients are also analysed.Keywords
This publication has 58 references indexed in Scilit:
- Lung tumor motion measured using retrospective 4D CT and correlated with tumor location and attachmentInternational Journal of Radiation Oncology*Biology*Physics, 2004
- Reproducibility of lung tumor position and reduction of lung mass within the planning target volume using active breathing control (ABC)International Journal of Radiation Oncology*Biology*Physics, 2003
- Respiration‐correlated spiral CT: A method of measuring respiratory‐induced anatomic motion for radiation treatment planningMedical Physics, 2002
- Correcting organ motion artifacts in x‐ray CT systems based on tracking of motion phase by the spatial overlap correlator. II. Experimental studyMedical Physics, 2001
- Correcting organ motion artifacts in x‐ray CT medical imaging systems by adaptive processing. I. TheoryMedical Physics, 2001
- Noninvasive Coronary Angiography by Retrospectively ECG-Gated Multislice Spiral CTCirculation, 2000
- Breathing pattern in humans: diversity and individualityRespiration Physiology, 2000
- What margins should be added to the clinical target volume in radiotherapy treatment planning for lung cancer?Radiotherapy and Oncology, 1998
- Target volume definition for three-dimensional conformal radiation therapy of lung cancer.The British Journal of Radiology, 1998
- Improved survival of patients with unresectable non-small-cell bronchogenic carcinoma by an innovated high-doseEn-Bloc radiotherapeutic approachCancer, 1981