Blurring and Layer Thickness in Narrow Beam Rotation Radiography
- 1 November 1977
- journal article
- Published by SAGE Publications in Acta Radiologica. Diagnosis
- Vol. 18 (6) , 705-714
- https://doi.org/10.1177/028418517701800615
Abstract
The definition of the blurring of narrow beam rotation radiography is revived. The relative value of the blurring is considered to be the most reliable measurement. Calculations of the layer thickness should be based on the definition of the relative blurring. Experimental tests were performed, which confirm the mathematically presented hypothesis and also confirm the validity of the mathematical model of narrow beam rotation radiography presented by WELANDER (1974).Keywords
This publication has 4 references indexed in Scilit:
- Layer Formation in Narrow Beam Rotation RadiographyActa Radiologica. Diagnosis, 1975
- A Mathematical Model of Narrow Beam Rotation MethodsActa Radiologica. Diagnosis, 1974
- Pantomography in Theory and UseActa Radiologica, 1954
- A New Tomographical Method for Radiographing Curved Outer SurfacesActa Radiologica, 1949