Agilent 2100 Bioanalyzer for Restriction Fragment Length Polymorphism Analysis of the Campylobacter jejuni Flagellin Gene
Open Access
- 1 February 2001
- journal article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 39 (2) , 754-757
- https://doi.org/10.1128/jcm.39.2.754-757.2001
Abstract
The Agilent 2100 bioanalyzer (Agilent Technologies, Palo Alto, Calif.) utilizes capillary electrophoresis on a microchip device (LabChip 7500; Caliper Technologies, Mountain View, Calif.) that is capable of rapidly sizing small DNA fragments. To determine whether the system could replace conventional restriction fragment length polymorphism (RFLP) typing by agarose gel electrophoresis, we compared the analyzer with conventional flagellin RFLP for typingCampylobacter jejuni. Ninety-seven isolates representing 46 Fla types were initially analyzed. Correct Fla types were detected in 59% of the isolates. The major problem with the system was in resolving samples containing multiple DNA fragments differing from 8 to 20 bp. Overall, the bioanalyzer has the potential to replace conventional RFLP analysis by gel electrophoresis, but improvements in the chip separation are needed.Keywords
This publication has 8 references indexed in Scilit:
- Evaluation of DNA Fragment Sizing and Quantification by the Agilent 2100 BioanalyzerClinical Chemistry, 2000
- Genotyping of Campylobacter sppApplied and Environmental Microbiology, 2000
- Safer and healthier foods.1999
- Quinolone-ResistantCampylobacter jejuniInfections in Minnesota, 1992–1998New England Journal of Medicine, 1999
- Analysis of HL and O serotypes of Campylobacter strains by the flagellin gene typing systemJournal of Clinical Microbiology, 1996
- Flagellin gene typing of Campylobacter jejuni by restriction fragment length polymorphism analysisJournal of Clinical Microbiology, 1993