Magnetic resonance imaging of microbubbles in a superheated emulsion chamber for brachytherapy dosimetry
- 3 December 1998
- journal article
- Published by Wiley in Medical Physics
- Vol. 25 (12) , 2316-2325
- https://doi.org/10.1118/1.598441
Abstract
This paper describes development of magnetic resonance imaging (MRI) techniques for three‐dimensional (3D) imaging of a position‐sensitive detector for brachytherapy dosimetry. The detector is a 0.5 l chamber containing an emulsion of halocarbon‐115 droplets in a tissue‐equivalent glycerin‐based gel. The halocarbon droplets are highly superheated and expand into vapor microbubbles upon irradiation. Brachytherapy sources can be inserted into the superheated emulsion chamber to create distributions of bubbles. Three‐dimensional MRI of the chamber is then performed. A 3D gradient‐echo technique was optimized for spatial resolution and contrast between bubbles and gel. Susceptibility gradients at the interfaces between bubbles and gel are exploited to enhance contrast so microscopic bubbles can be imaged using relatively large voxel sizes. Three‐dimensional gradient‐echo images are obtained with an isotropic resolution of 300 μm over a field‐of‐view in an imaging time of 14 min. A post‐processing technique was developed to semi‐automatically segment the bubbles from the images and to assess dose distributions based on the measured bubble densities. Relative dose distributions are computed from MR images for a brachytherapy source and the results compare favorably to relative radial dose distributions calculated as recommended by Task Group 43 of the American Association of Physicists in Medicine.Keywords
This publication has 21 references indexed in Scilit:
- Implementation of wavelet‐encoded MR imagingJournal of Magnetic Resonance Imaging, 1993
- Dose distributions produced by shielded applicators using 241 Am for intracavitary irradiation of tumors in the vaginaMedical Physics, 1992
- Wavelet‐encoded MR imagingMagnetic Resonance in Medicine, 1992
- Reduction of susceptibility artifact in gradient‐echo imagingMagnetic Resonance in Medicine, 1992
- MR fluoroscopy: Technical feasibilityMagnetic Resonance in Medicine, 1988
- Dose distribution for 1 2 5 I implants due to anisotropic radiation emission and unknown seed orientationMedical Physics, 1987
- Susceptibility artefacts in NMR imagingMagnetic Resonance Imaging, 1985
- Measurement of radiation dose distributions by nuclear magnetic resonance (NMR) imagingPhysics in Medicine & Biology, 1984
- Dose inhomogeneity in interstitial implants using 125I seedsInternational Journal of Radiation Oncology*Biology*Physics, 1979
- Some Effects of Ionizing Radiation on the Formation of Bubbles in LiquidsPhysical Review B, 1952