A novel, open access, elliptical cross-section magnet for paediatric MRI
- 1 January 1998
- journal article
- Published by IOP Publishing in Measurement Science and Technology
- Vol. 9 (1) , 113-119
- https://doi.org/10.1088/0957-0233/9/1/015
Abstract
Almost all clinical magnetic resonance imaging systems are based on circular cross-section magnets. Recent advances in elliptical cross-section RF probe and gradient coil hardware raise the question of the possibility of using elliptical cross-section magnet systems. This paper presents a methodology for calculating rapidly the magnetic fields generated by a multi-turn coil of elliptical cross-section and incorporates this in a stochastic optimization method for magnet design. An open magnet system of elliptical cross-section is designed that both reduces the claustrophobia for the patients and allows ready access by attending physicians. The magnet system is optimized for paediatric use. The coil geometry produced by the optimization method has several novel features.Keywords
This publication has 12 references indexed in Scilit:
- Rapid computation of static fields produced by thick circular solenoidsIEEE Transactions on Magnetics, 1997
- General Expressions for the Magnetic Field and Stored Energy of Elliptic Cylinder CoilsJournal of Magnetic Resonance, Series B, 1996
- Improved, Minimum-Inductance, Elliptic-Cylinderz-Gradient Coil Using Axial and Azimuthal Current FlowJournal of Magnetic Resonance, Series B, 1996
- Optimization of Transverse Gradient Coils with Coaxial Return Paths by Simulated AnnealingJournal of Magnetic Resonance, Series B, 1996
- An analysis and optimization of elliptical RF probes used in magnetic resonance imagingMeasurement Science and Technology, 1996
- Superconducting open-configuration MR imaging system for image-guided therapy.Radiology, 1995
- The Design of Transverse Gradient Coils of Restricted Length by Simulated AnnealingJournal of Magnetic Resonance, Series A, 1994
- Gradient-Coil Design by Simulated AnnealingJournal of Magnetic Resonance, Series A, 1993
- An MRI elliptical coil with minimum inductanceMeasurement Science and Technology, 1993
- Considerations in the design of MRI magnets with reduced stray fieldsIEEE Transactions on Magnetics, 1987