Contact force and force-time integral in atrial radiofrequency ablation predict transmurality of lesions
Open Access
- 1 May 2014
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 16 (5) , 660-667
- https://doi.org/10.1093/europace/euu068
Abstract
During radiofrequency (RF) delivery, lesion volume is highly dependent on contact force (CF). It has recently been shown that changes of bipolar electrogram (EGM) predict transmurality. We hypothesized that there is a correlation between CF and EGM criteria of transmural lesion (TL) during RF. We prospectively studied consecutive 512 RF applications from atrial fibrillation ablation procedures. A force-sensing ablation catheter (Tacticath®, Endosense) was used to continuously measure CF and force-time integral (FTI) during each RF application. Distal bipolar EGM was analysed before, during, and after each RF application. Depending on initial EGM morphology, transmurality of lesions was defined by: (i) disappearance of the positivity after RF when there was QR morphology, (ii) diminution >75% of the positivity when there was QRS morphology, or (iii) disappearance of the R′ positivity when there was RSR′ morphology. Electrogram criteria were validated by electrophysiologists blinded to force measurements. Force-time integral was higher in TL than in non-transmural lesions (NTLs): 652 ± 248 vs. 212 ± 140 gs (P < 0.001). Mean CF per RF pulse was higher in TL than in NTL: 26.3 ± 12.5 vs. 11.3 ± 10.3 g (P < 0.001). The best cut-off to predict TL was an FTI ≥ 392 gs [sensitivity 0.89, specificity (Sp) 0.93, positive predictive value (PPV) 0.98, and negative predictive value 0.67] and a higher FTI (>700 gs) warrants transmurality of RF atrial lesions (100% Sp and PPV). Contact force and FTI during RF are correlated with TL. During RF delivery, a target FTI > 392 gs can be used as an endpoint.Keywords
This publication has 16 references indexed in Scilit:
- 2012 HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Patient Selection, Procedural Techniques, Patient Management and Follow-up, Definitions, Endpoints, and Research Trial DesignA report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation. Developed in partnership with the European Heart Rhythm Association (EHRA), a registered branch of the European Society of Cardiology (ESC) and the European Cardiac Arrhythmia Society (ECAS); and in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), the Asia Pacific Heart Rhythm Society (APHRS), and the Society of Thoracic Surgeons (STS). Endorsed by the governing bodies of the American College of Cardiology Foundation, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, the Asia Pacific Heart Rhythm Society, and the Heart Rhythm SocietyEP Europace, 2012
- A novel radiofrequency ablation catheter using contact force sensing: Toccata studyHeart Rhythm, 2012
- Local unipolar and bipolar electrogram criteria for evaluating the transmurality of atrial ablation lesions at different catheter orientations relative to the endocardial surfaceHeart Rhythm, 2010
- Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and ThrombusCirculation: Arrhythmia and Electrophysiology, 2008
- Electrophysiological Findings During Ablation of Persistent Atrial Fibrillation With Electroanatomic Mapping and Double Lasso Catheter TechniqueCirculation, 2005
- Late recurrent arrhythmias after ablation of atrial fibrillation: Incidence, mechanisms, and treatmentHeart Rhythm, 2004
- Simultaneous creation and evaluation of linear radiofrequency lesions.Journal of Interventional Cardiac Electrophysiology, 2002
- Electrogram-Guided Radiofrequency Catheter Ablation of Atrial Tissue Comparison with Thermometry-Guide AblationJournal of Interventional Cardiac Electrophysiology, 2001
- Atrial Linear Ablations in PigsCirculation, 1999
- Monitoring the Local Electrogram at the Ablation Site during Radiofrequency Application for Common Atrial FlutterJapanese Circulation Journal, 1998