Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus
Top Cited Papers
Open Access
- 1 December 2008
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation: Arrhythmia and Electrophysiology
- Vol. 1 (5) , 354-362
- https://doi.org/10.1161/circep.108.803650
Abstract
Background— An open-irrigated radiofrequency (RF) ablation catheter was developed to measure contact force (CF). Three optical fibers measure microdeformation of the catheter tip. The purpose of this study was to (1) validate the accuracy of CF sensor (CFS) (bench test); and (2) determine the relationship between CF and tissue temperatures, lesion size, steam pop, and thrombus during RF ablation using a canine thigh muscle preparation.Keywords
This publication has 17 references indexed in Scilit:
- RF Catheter Ablation: Lessons on LesionsPacing and Clinical Electrophysiology, 2006
- Comparison of Electrode Cooling Between Internal and Open Irrigation in Radiofrequency Ablation Lesion Depth and Incidence of Thrombus and Steam PopCirculation, 2006
- Saline‐Irrigated Radiofrequency Ablation Electrode with External CoolingJournal of Cardiovascular Electrophysiology, 2005
- Tissue Temperature‐Controlled Radiofrequency AblationPacing and Clinical Electrophysiology, 2003
- Efficacy and safety of septal and left-atrial linear ablation for atrial fibrillationThe American Journal of Cardiology, 1999
- The LETR‐Principle: A Novel Method to Assess Electrode‐Tissue Contact in Radiofrequency AblationJournal of Cardiovascular Electrophysiology, 1998
- Effects of dispersive electrode position and surface area on electrical parameters and temperature during radiofrequency catheter ablationThe American Journal of Cardiology, 1996
- Embolic complications associated with radiofrequency catheter ablationThe American Journal of Cardiology, 1996
- Embolic complications after radiofrequency catheter ablationThe American Journal of Cardiology, 1994
- Determinants of Lesion Size During Radiofrequency Catheter Ablation: The Role of Electrode‐Tissue Contact Pressure and Duration of Energy DeliveryJournal of Cardiovascular Electrophysiology, 1991