Closed-loop Control of Blood Glucose Levels in Critically Ill Patients
Open Access
- 1 June 2002
- journal article
- research article
- Published by SAGE Publications in Anaesthesia and Intensive Care
- Vol. 30 (3) , 295-307
- https://doi.org/10.1177/0310057x0203000306
Abstract
A closed-loop control system was constructed for automatic intravenous infusion of insulin to control blood sugar levels (BSL) in critically ill patients. We describe the development of the system. A total of nine subjects were recruited to clinically test the control system. In the patients who underwent closed-loop control of BSL, the controller managed to control only one patient's glycaemia without any manual intervention. The average BSL attained during closed-loop control approached the target range of 6-10 mmol/l, and had less deviation than when BSL had been maintained manually. We conclude that closed-loop BSL control using a sliding scale algorithm is feasible. The main deficiency in the current system is unreliability of the subcutaneous glucose sensor when used in this setting. This deficiency mandates high vigilance during use of the system as it is being developed.Keywords
This publication has 19 references indexed in Scilit:
- Intensive Insulin Therapy in Critically Ill PatientsNew England Journal of Medicine, 2001
- Glucose Clamps with the Biostator: A Critical ReappraisalHormone and Metabolic Research, 1994
- Development of a Miniaturized Glucose Monitoring System by Combining a Needle-Type Glucose Sensor With Microdialysis Sampling Method: Long-term subcutaneous tissue glucose monitoring in ambulatory diabetic patientsDiabetes Care, 1994
- A glucose monitoring system for on line estimation in man of blood glucose concentration using a miniaturized glucose sensor implanted in the subcutaneous tissue and a wearable control unitDiabetologia, 1993
- Devices for Insulin AdministrationDiabetes Care, 1990
- Automated feedback control of subcutaneous glucose concentration in diabetic dogsDiabetologia, 1989
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987
- Glycaemic control in pancreatectomized dogs with a wearable artificial endocrine pancreasDiabetologia, 1983