Ability of the Bispectral Index, Autoregressive Modelling with Exogenous Input-derived Auditory Evoked Potentials, and Predicted Propofol Concentrations to Measure Patient Responsiveness during Anesthesia with Propofol and Remifentanil
- 1 October 2003
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Anesthesiology
- Vol. 99 (4) , 802-812
- https://doi.org/10.1097/00000542-200310000-00010
Abstract
Background: This study was conducted to compare the performance accuracy of the independent variables Bispectral Index (BIS), A-Line ARX index (AAI), and predicted propofol effect-site concentration (CePROP) to measure the dependent variables of loss of responses to different stimulation defined as loss of response to verbal command (LORverbal), eyelash reflex (LORlash), and noxious stimulus (LORnoxious) during stepwise increased levels of propofol infusion with and without remifentanil. Methods: Forty-five patients were randomly allocated to one of three groups (0, 2, and 4 ng/ml remifentanil) to receive graded CePROP and predicted effect compartment controlled remifentanil (CeREMI). At every step, the ability to respond to verbal command using the Observer's Assessment of Alertness/Sedation Scale (OAA/S), eyelash reflex, and electrical tetanic noxious stimulus were compared against BIS, AAI, and CePROP. Prediction probability and sensitivity/specificity were calculated. Results: Increasing CeREMI increased BIS and AAI values at LORverbal and LORlash and decreased CePROP. Similar findings were found for LORnoxious. The overall prediction probability to measure the hypnotic component of anesthesia remained accurate in the three groups for BIS, AAI, and CePROP. Combined information from CePROP, CeREMI, and BIS or AAI increased the overall prediction probability for predicting the OAA/S scale and LORlash. Less accuracy to LORnoxious was found in all independent variables. Conclusions: Although BIS, AAI, and CePROP were influenced by remifentanil during propofol administration, their ability to detect OAA/S and LORlash remained accurate. Improved performance is obtained when BIS and AAI are measured in conjunction with drug targeted effect-site concentrations. Remifentanil decreases the ability of these independent variables to detect LORnoxious.Keywords
This publication has 35 references indexed in Scilit:
- Performance of the ARX-derived Auditory Evoked Potential Index as an Indicator of Anesthetic DepthAnesthesiology, 2002
- Comparison of auditory evoked potentials and the A‐line ARX Index for monitoring the hypnotic level during sevoflurane and propofol inductionActa Anaesthesiologica Scandinavica, 2002
- Effect of sevoflurane on the mid-latency auditory evoked potentials measured by a new fast extracting monitorActa Anaesthesiologica Scandinavica, 2002
- Shannon Entropy Applied to the Measurement of the Electroencephalographic Effects of DesfluraneAnesthesiology, 2001
- Monitoring Depth of AnesthesiaAnesthesiology, 2000
- Changes in rapidly extracted auditory evoked potentials during tracheal intubationActa Anaesthesiologica Scandinavica, 2000
- Approximate Entropy as an Electroencephalographic Measure of Anesthetic Drug Effect during Desflurane AnesthesiaAnesthesiology, 2000
- Comparison of bispectral index, 95% spectral edge frequency and approximate entropy of the EEG, with changes in heart rate variability during induction of general anaesthesiaBritish Journal of Anaesthesia, 1999
- On-line analysis of middle latency auditory evoked potentials (MLAEP) for monitoring depth of anaesthesia in laboratory ratsMedical Engineering & Physics, 1999
- Evoked responses in anaesthesiaBritish Journal of Anaesthesia, 1998