Monte Carlo verification of IMRT dose distributions from a commercial treatment planning optimization system
- 25 August 2000
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 45 (9) , 2483-2495
- https://doi.org/10.1088/0031-9155/45/9/303
Abstract
The purpose of this work was to use Monte Carlo simulations to verify the accuracy of the dose distributions from a commercial treatment planning optimization system (Corvus, Nomos Corp., Sewickley, PA) for intensity-modulated radiotherapy (IMRT). A Monte Carlo treatment planning system has been implemented clinically to improve and verify the accuracy of radiotherapy dose calculations. Further modifications to the system were made to compute the dose in a patient for multiple fixed-gantry IMRT fields. The dose distributions in the experimental phantoms and in the patients were calculated and used to verify the optimized treatment plans generated by the Corvus system. The Monte Carlo calculated IMRT dose distributions agreed with the measurements to within 2% of the maximum dose for all the beam energies and field sizes for both the homogeneous and heterogeneous phantoms. The dose distributions predicted by the Corvus system, which employs a finite-size pencil beam (FSPB) algorithm, agreed with the Monte Carlo simulations and measurements to within 4% in a cylindrical water phantom with various hypothetical target shapes. Discrepancies of more than 5% (relative to the prescribed target dose) in the target region and over 20% in the critical structures were found in some IMRT patient calculations. The FSPB algorithm as implemented in the Corvus system is adequate for homogeneous phantoms (such as prostate) but may result in significant under- or over-estimation of the dose in some cases involving heterogeneities such as the air-tissue, lung-tissue and tissue-bone interfaces.Keywords
This publication has 28 references indexed in Scilit:
- Theoretical considerations of monitor unit calculations for intensity modulated beam treatment planningMedical Physics, 1999
- Monte Carlo calculations of electron beam output factors for a medical linear acceleratorPhysics in Medicine & Biology, 1998
- Monitor chamber backscatter for intensity modulated radiation therapy using multileaf collimatorsPhysics in Medicine & Biology, 1998
- A CT‐based Monte Carlo simulation tool for dosimetry planning and analysisMedical Physics, 1998
- Modulated beam conformal therapy for head and neck tumorsInternational Journal of Radiation Oncology*Biology*Physics, 1997
- Three dimensional conformal treatment planning with multileaf collimatorsInternational Journal of Radiation Oncology*Biology*Physics, 1995
- A computer-controlled conformal radiotherapy system I: OverviewInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Dose calculation for photon beams with intensity modulation generated by dynamic jaw or multileaf collimationsMedical Physics, 1994
- The generation of intensity-modulated fields for conformal radiotherapy by dynamic collimationPhysics in Medicine & Biology, 1992
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983