Central brightening due to constructive interference with, without, and despite dielectric resonance
Top Cited Papers
Open Access
- 21 January 2005
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 21 (2) , 192-196
- https://doi.org/10.1002/jmri.20245
Abstract
Purpose To aid in discussion about the mechanism for central brightening in high field magnetic resonance imaging (MRI), especially regarding the appropriateness of using the term dielectric resonance to describe the central brightening seen in images of the human head. Materials and Methods We present both numerical calculations and experimental images at 3 T of a 35-cm-diameter spherical phantom of varying salinity both with one surface coil and with two surface coils on opposite sides, and further numerical calculations at frequencies corresponding to dielectric resonances for the sphere. Results With two strategically placed surface coils it is possible to create central brightening even when one coil alone excites an image intensity pattern either bright on one side only or bright on both sides with central darkening. This central brightening can be created with strategic coil placement even when the resonant pattern would favor central darkening. Results in a conductive sample show that central brightening can similarly be achieved in weakly conductive dielectric materials where any true resonances would be heavily damped, such as in human tissues. Conclusion Constructive interference and wavelength effects are likely bigger contributors to central brightening in MR images of weakly conductive biological samples than is true dielectric resonance. J. Magn. Reson. Imaging 2005;21:192–196.Keywords
This publication has 17 references indexed in Scilit:
- Image brightening in samples of high dielectric constantJournal of Magnetic Resonance, 2004
- Analysis of wave behavior in lossy dielectric samples at high fieldMagnetic Resonance in Medicine, 2002
- Different excitation and reception distributions with a single‐loop transmit‐receive surface coil near a head‐sized spherical phantom at 300 MHzMagnetic Resonance in Medicine, 2002
- Radio frequency magnetic field mapping of a 3 Tesla birdcage coil: Experimental and theoretical dependence on sample propertiesMagnetic Resonance in Medicine, 2001
- Signal‐to‐noise ratio and absorbed power as functions of main magnetic field strength, and definition of “90°” RF pulse for the head in the birdcage coilMagnetic Resonance in Medicine, 2001
- Standing-wave and RF penetration artifacts caused by elliptic geometry: an electrodynamic analysis of MRIIEEE Transactions on Medical Imaging, 1998
- Standing Waves in Uniform Water PhantomsJournal of Magnetic Resonance, Series B, 1994
- In vivo magnetic resonance imaging and spectroscopy of humans with a 4 t whole‐body magnetNMR in Biomedicine, 1990
- Spectroscopy and imaging with a 4 tesla whole‐body mr systemNMR in Biomedicine, 1988